Purpose: To report an alternative technique to the dual-lumen catheter for deployment of the Powerlink stent-graft in patients with angulated sacs and calcified aortic bifurcations. A maneuver is also presented to retrieve the delivery system when it is snagged on the stent. Technique: After cutdown of the right common femoral artery (CFA), a 9-F introducer sheath is placed percutaneously into the left CFA. A gooseneck catheter is introduced from the right CFA to capture a 0.035-inch hydrophilic guidewire inserted from the left. A 5-F straight catheter is passed over this guidewire from the left to the right CFA. In angulated aneurysm sacs, a 5-F Hunter catheter is introduced from the right femoral access to support a guidewire through the aneurysm to the suprarenal aorta. Then the guidewire is exchanged with a 0.035-inch Amplatz extra stiff wire, and the Hunter catheter is removed. In other cases, a 0.035-inch Amplatz extra stiff guidewire is placed up to the suprarenal aorta. The endograft delivery system is then deployed in the usual manner. A gooseneck snare is also useful in retrieving the delivery system when it is snagged on the stent at the endograft bifurcation. Conclusions: This variant technique facilitates the deployment of the Powerlink stent-graft when faced with angulated aneurysms or acute and calcified aortic bifurcations. A goose-neck catheter is helpful in retrieving the delivery system's "olive" after endograft placement.

A variant deployment technique for the Powerlink bifurcated endograft / G. Nano,I. Dalainas, P. Volpe, R. Casana, T. Lupatelli, G. Paroni, L. Inglese, D.G. Tealdi. - In: JOURNAL OF ENDOVASCULAR THERAPY. - ISSN 1526-6028. - 12:6(2005), pp. 638-641.

A variant deployment technique for the Powerlink bifurcated endograft

G. Nano
Primo
;
D.G. Tealdi
Ultimo
2005

Abstract

Purpose: To report an alternative technique to the dual-lumen catheter for deployment of the Powerlink stent-graft in patients with angulated sacs and calcified aortic bifurcations. A maneuver is also presented to retrieve the delivery system when it is snagged on the stent. Technique: After cutdown of the right common femoral artery (CFA), a 9-F introducer sheath is placed percutaneously into the left CFA. A gooseneck catheter is introduced from the right CFA to capture a 0.035-inch hydrophilic guidewire inserted from the left. A 5-F straight catheter is passed over this guidewire from the left to the right CFA. In angulated aneurysm sacs, a 5-F Hunter catheter is introduced from the right femoral access to support a guidewire through the aneurysm to the suprarenal aorta. Then the guidewire is exchanged with a 0.035-inch Amplatz extra stiff wire, and the Hunter catheter is removed. In other cases, a 0.035-inch Amplatz extra stiff guidewire is placed up to the suprarenal aorta. The endograft delivery system is then deployed in the usual manner. A gooseneck snare is also useful in retrieving the delivery system when it is snagged on the stent at the endograft bifurcation. Conclusions: This variant technique facilitates the deployment of the Powerlink stent-graft when faced with angulated aneurysms or acute and calcified aortic bifurcations. A goose-neck catheter is helpful in retrieving the delivery system's "olive" after endograft placement.
Abdominal aortic aneurysm; Delivery technique; Endovascular repair; Stent-graft
Settore MED/22 - Chirurgia Vascolare
2005
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/13888
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
social impact