Purpose To characterize the image quality and dose level of a cone-beam CT (CBCT) scanner dedicated to maxillo-facial intra-surgical navigation. Methods Acquisition protocols for head imaging of average pediatric and adult patients were investigated. The detector response curve and the modulation transfer function (MTF) were assessed. The 3D image quality was evaluated using a commercial Catphan604 phantom in terms of accuracy of Hounsfield Unit (HU) estimates and slice uniformity. The MTF and the noise power spectrum (NPS) were assessed over the 3D reconstructed images. The weighted cone-beam dose index (CBDIw) was used for dosimetric benchmarking, and TLD chips placed in two anthropomorphic phantoms provided insight into the patient dose distribution. Results The detector exhibited a response curve with a double natural logarithmic behavior and spatial resolutions up to 6.3 mm−1. The 3D spatial resolution ranged from 0.70 mm−1 at the scanner isocenter, for the pediatric protocol, to 0.46 mm−1 at 7 cm from the isocenter for the adult protocol. Due to beam hardening and the scatter, discrepancies between expected and measured HU were observed, in particular for positive HU values (i.e. those for acrylic, bone, delrin and teflon inserts). The CBDIw and dose distribution within anthropomorphic phantoms provided a dosimetric benchmark for the scanner. Conclusion This study characterized a CBCT scanner in maxillo-facial surgery navigation. The findings support the definition of quality assurance protocols, providing methodologies and reference values for image quality and dose, while also highlighting limitations in spatial resolution and accuracy relevant for quantitative evaluations in clinical practice.

Physical characterization of a CBCT scanner for maxillo-facial intra-surgical navigation / C. Cotena, C.M.. - In: PHYSICA MEDICA. - ISSN 1120-1797. - 147:(2026 Jul), pp. 105818.1-105818.11. [10.1016/j.ejmp.2026.105818]

Physical characterization of a CBCT scanner for maxillo-facial intra-surgical navigation

A. Sarno
Ultimo
2026

Abstract

Purpose To characterize the image quality and dose level of a cone-beam CT (CBCT) scanner dedicated to maxillo-facial intra-surgical navigation. Methods Acquisition protocols for head imaging of average pediatric and adult patients were investigated. The detector response curve and the modulation transfer function (MTF) were assessed. The 3D image quality was evaluated using a commercial Catphan604 phantom in terms of accuracy of Hounsfield Unit (HU) estimates and slice uniformity. The MTF and the noise power spectrum (NPS) were assessed over the 3D reconstructed images. The weighted cone-beam dose index (CBDIw) was used for dosimetric benchmarking, and TLD chips placed in two anthropomorphic phantoms provided insight into the patient dose distribution. Results The detector exhibited a response curve with a double natural logarithmic behavior and spatial resolutions up to 6.3 mm−1. The 3D spatial resolution ranged from 0.70 mm−1 at the scanner isocenter, for the pediatric protocol, to 0.46 mm−1 at 7 cm from the isocenter for the adult protocol. Due to beam hardening and the scatter, discrepancies between expected and measured HU were observed, in particular for positive HU values (i.e. those for acrylic, bone, delrin and teflon inserts). The CBDIw and dose distribution within anthropomorphic phantoms provided a dosimetric benchmark for the scanner. Conclusion This study characterized a CBCT scanner in maxillo-facial surgery navigation. The findings support the definition of quality assurance protocols, providing methodologies and reference values for image quality and dose, while also highlighting limitations in spatial resolution and accuracy relevant for quantitative evaluations in clinical practice.
Settore PHYS-06/A - Fisica per le scienze della vita, l'ambiente e i beni culturali
   Non-circular trajectory and collimated beam towards quantitative Cone-beam Computed Tomography” (Q – CT)
   Q - CT
   MINISTERO DELL'UNIVERSITA' E DELLA RICERCA
   2022EHXL8M
lug-2026
Article (author)
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S1120179726001055-main.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Licenza: Creative commons
Dimensione 6.41 MB
Formato Adobe PDF
6.41 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1275760
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
  • OpenAlex ND
social impact