Laboratory-grade force platforms are the reference standard for quantifying ground reaction forces but remain costly and confined to well-resourced laboratories, limiting field-based use. This study evaluated the accuracy and between-session reliability of a portable force platform (Impera, Spinitalia) under static loading, and its concurrent validity during vertical jumping, against a criterion-standard piezoelectric plate with calibration traceable to national standards (4Jump, Kistler). Accuracy and between-session reliability were assessed against five tared loads (21.6–103.2 kg) applied at five plate locations across two sessions 24 h apart. Concurrent validity was assessed during squat and countermovement jumps performed by ten physically active men, using intraclass correlation coefficients, Bland–Altman analysis and the linear fit method applied to the entire force–time waveform. Both platforms underestimated the lightest load by about 4% and approached zero error at higher loads; the between-device difference was 0.19 kg and did not translate into a difference in proportional accuracy (p = 0.151), with no device × position interaction. Between-session coefficients of variation were below 1% for both devices. Concurrent agreement was good to excellent for all jump variables (ICC 0.867–0.996), and waveform agreement was close to unity (R2 ≥ 0.997), with about 2% amplitude underestimation, greater in participants generating the highest impact forces. The Impera platform provides accurate and reliable measurements under static loading, and force–time data that agree closely with the 4Jump during vertical jump testing.
Accuracy, Reliability and Waveform-Level Agreement Between the Impera Force Platform and a Criterion-Standard Force Plate / D. Pompa, A.C.. - In: SENSORS. - ISSN 1424-8220. - 26:18(2026 Sep 02), pp. 5850.1-5850.24. [10.3390/s26185850]
Accuracy, Reliability and Waveform-Level Agreement Between the Impera Force Platform and a Criterion-Standard Force Plate
J. PaduloUltimo
2026
Abstract
Laboratory-grade force platforms are the reference standard for quantifying ground reaction forces but remain costly and confined to well-resourced laboratories, limiting field-based use. This study evaluated the accuracy and between-session reliability of a portable force platform (Impera, Spinitalia) under static loading, and its concurrent validity during vertical jumping, against a criterion-standard piezoelectric plate with calibration traceable to national standards (4Jump, Kistler). Accuracy and between-session reliability were assessed against five tared loads (21.6–103.2 kg) applied at five plate locations across two sessions 24 h apart. Concurrent validity was assessed during squat and countermovement jumps performed by ten physically active men, using intraclass correlation coefficients, Bland–Altman analysis and the linear fit method applied to the entire force–time waveform. Both platforms underestimated the lightest load by about 4% and approached zero error at higher loads; the between-device difference was 0.19 kg and did not translate into a difference in proportional accuracy (p = 0.151), with no device × position interaction. Between-session coefficients of variation were below 1% for both devices. Concurrent agreement was good to excellent for all jump variables (ICC 0.867–0.996), and waveform agreement was close to unity (R2 ≥ 0.997), with about 2% amplitude underestimation, greater in participants generating the highest impact forces. The Impera platform provides accurate and reliable measurements under static loading, and force–time data that agree closely with the 4Jump during vertical jump testing.| File | Dimensione | Formato | |
|---|---|---|---|
|
Pompa 2026 Sensors Accuracy, Reliability and Waveform-Level Agreement Between the Impera Force Platform and a Criterion-Standard Force Plate.pdf
accesso aperto
Tipologia:
Publisher's version/PDF
Licenza:
Creative commons
Dimensione
1.67 MB
Formato
Adobe PDF
|
1.67 MB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.