(1) Background: Preclinical and clinical data about a novel radiofrequency ablation (RFA) system (STARmed Co, Ltd.; Koyang, Korea) designed to be used under endoscopic ultrasound (EUS) control for pancreatic lesion ablation, are limited, obtained with non-standardized procedures and heterogeneous results. The aim of this study is to standardize the RFA procedure of this system in order to define the optimal ablation power and time. (2) Methods: RFA was performed on an ex-vivo porcine liver at different powers (40, 30, 20, 10 Watts (W)) and times (1, 3, 5, 7, 15 min) with a 1-centimeter monopolar electrode (perfused by chilled solution) positioned on the distal tip of a 19-Gauge needle. A blinded expert pathologist histologically analyzed each ablation area. (3) Results: The size of the total macroscopic ablated area was negatively correlated with ablation power (R −0.74): the largest was obtained at 10 W (p = 4.7 × 10−4) for longer times (R 0.92; p = 8.9 × 10−8). Central histologic coagulative necrosis did not differ among ablation settings (mean size 3.25 mm). External “parenchymal hypochromia” or “diaphanization” resulted the widest at 10 W, for longer times (R 0.8, p = 3.6 × 10−4). (4) Conclusions: The RFA system can produce small sizes of coagulative necrosis, regardless of the setting. Larger areas of diaphanization surrounding the necrosis can be produced at lower powers for longer times.

Standardization of a Radiofrequency Ablation Tool in an Ex-Vivo Porcine Liver Model / G. Rossi, M.C. Petrone, G. Capurso, L. Albarello, S.G.G. Testoni, L. Archibugi, M. Schiavo Lena, C. Doglioni, P.G. Arcidiacono. - In: GASTROINTESTINAL DISORDERS. - ISSN 2624-5647. - 2:3(2020 Sep), pp. 27.300-27.309. [10.3390/gidisord2030027]

Standardization of a Radiofrequency Ablation Tool in an Ex-Vivo Porcine Liver Model

S.G.G. Testoni;
2020

Abstract

(1) Background: Preclinical and clinical data about a novel radiofrequency ablation (RFA) system (STARmed Co, Ltd.; Koyang, Korea) designed to be used under endoscopic ultrasound (EUS) control for pancreatic lesion ablation, are limited, obtained with non-standardized procedures and heterogeneous results. The aim of this study is to standardize the RFA procedure of this system in order to define the optimal ablation power and time. (2) Methods: RFA was performed on an ex-vivo porcine liver at different powers (40, 30, 20, 10 Watts (W)) and times (1, 3, 5, 7, 15 min) with a 1-centimeter monopolar electrode (perfused by chilled solution) positioned on the distal tip of a 19-Gauge needle. A blinded expert pathologist histologically analyzed each ablation area. (3) Results: The size of the total macroscopic ablated area was negatively correlated with ablation power (R −0.74): the largest was obtained at 10 W (p = 4.7 × 10−4) for longer times (R 0.92; p = 8.9 × 10−8). Central histologic coagulative necrosis did not differ among ablation settings (mean size 3.25 mm). External “parenchymal hypochromia” or “diaphanization” resulted the widest at 10 W, for longer times (R 0.8, p = 3.6 × 10−4). (4) Conclusions: The RFA system can produce small sizes of coagulative necrosis, regardless of the setting. Larger areas of diaphanization surrounding the necrosis can be produced at lower powers for longer times.
animal models; EUS; ex-vivo; pancreatic lesions; radiofrequency ablation
Settore MEDS-10/A - Gastroenterologia
set-2020
Article (author)
File in questo prodotto:
File Dimensione Formato  
gastrointestdisord-02-00027 Standardization of a Radiofrequency Ablation Tool 2020.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Licenza: Creative commons
Dimensione 2.89 MB
Formato Adobe PDF
2.89 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1229616
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 9
  • ???jsp.display-item.citation.isi??? ND
  • OpenAlex 10
social impact