The ratio of the nucleated cell counts (ΔTNC) provided by the Sysmex XT-2000iV DIFF and BASO channels has been proven to be useful in differentiating effusions in cats; in particular, a ΔTNC > 2.5 was found to have high accuracy for effusive FIP. The objective of this study was to compare the ΔTNC obtained with the Sysmex XT-2000iV and Sysmex XN-1000V on different types of effusions. Fresh effusions from cats were evaluated with both analyzers. WBC-BASO, WBC-DIFF, and ΔTNC-XT from the Sysmex XT-2000iV and TNC-WNR, TNC-WDF, and ΔTNC-XN from the Sysmex XN-1000V were recorded and compared. Forty-five feline effusions were collected. The results were not significantly different for all parameters evaluated. The Passing and Bablok regression analysis for ΔTNC-XT and ΔTNC-XN revealed a slope of 1.612 (95% CI 1.030 to 2.316) and an intercept of −0.744 (95% CI −1.591 to −0.120); the Bland–Altman difference plot showed a positive bias of 4.382 (p = 0.02). Despite the finding of a positive proportional and constant bias between ΔTNC-XT and ΔTNC-XN, evident especially at high ΔTNC values, samples were classified accordingly between the two analyzers when ΔTNC-XT cut-offs were used, allowing the same accuracy for the identification of suspected and compatible FIP samples.

Comparison of Delta Total Nucleated Cells Assessed via Sysmex XT-2000iV and Sysmex XN-1000V in Effusions in Cats / M. Zanetti, S. Meazzi, S. Paltrinieri, G. Mangiagalli, S. Novellini, S. Rossi, S. Klenner-Gastreich, S. Lauzi, A. Giordano. - In: ANIMALS. - ISSN 2076-2615. - 16:3(2026 Jan 23), pp. 366.1-366.10. [10.3390/ani16030366]

Comparison of Delta Total Nucleated Cells Assessed via Sysmex XT-2000iV and Sysmex XN-1000V in Effusions in Cats

S. Paltrinieri;S. Lauzi
Penultimo
;
A. Giordano
Ultimo
2026

Abstract

The ratio of the nucleated cell counts (ΔTNC) provided by the Sysmex XT-2000iV DIFF and BASO channels has been proven to be useful in differentiating effusions in cats; in particular, a ΔTNC > 2.5 was found to have high accuracy for effusive FIP. The objective of this study was to compare the ΔTNC obtained with the Sysmex XT-2000iV and Sysmex XN-1000V on different types of effusions. Fresh effusions from cats were evaluated with both analyzers. WBC-BASO, WBC-DIFF, and ΔTNC-XT from the Sysmex XT-2000iV and TNC-WNR, TNC-WDF, and ΔTNC-XN from the Sysmex XN-1000V were recorded and compared. Forty-five feline effusions were collected. The results were not significantly different for all parameters evaluated. The Passing and Bablok regression analysis for ΔTNC-XT and ΔTNC-XN revealed a slope of 1.612 (95% CI 1.030 to 2.316) and an intercept of −0.744 (95% CI −1.591 to −0.120); the Bland–Altman difference plot showed a positive bias of 4.382 (p = 0.02). Despite the finding of a positive proportional and constant bias between ΔTNC-XT and ΔTNC-XN, evident especially at high ΔTNC values, samples were classified accordingly between the two analyzers when ΔTNC-XT cut-offs were used, allowing the same accuracy for the identification of suspected and compatible FIP samples.
Delta-FIP; Sysmex XN-1000V; cats; cavitary effusions
Settore MVET-02/A - Patologia generale e anatomia patologica veterinaria
Settore MVET-03/A - Malattie infettive degli animali
23-gen-2026
Article (author)
File in questo prodotto:
File Dimensione Formato  
animals-16-00366.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Licenza: Creative commons
Dimensione 446.58 kB
Formato Adobe PDF
446.58 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1222457
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
  • OpenAlex ND
social impact