This work aims to provide the first characterization of the Loop-X Cone-Beam CT scanner for maxillofacial applications, with a focus on patient dose and image quality in 3D acquisitions. Measurements were performed with adult and paediatric “head” protocols, using the dedicated Catphan604 phantom to evaluate the different parameters. Spatial resolution was assessed using the modular transfer function, with a limiting spatial frequency —evaluated at 10% of its maximum— of 0.60 mm−1 for the adult protocol and 0.53 mm−1 for the paediatric protocol (mean value). In addition, noise was characterized in terms of noise power spectrum (NPS) and contrast to noise ratio (CNR) was calculated for the adult protocol between simulated soft tissue and bone (23 ± 1) and between spinal disc and soft tissue (1 ± 1).

3D image quality of a CBCT scanner for maxillofacial intrasurgical navigation / C. Cotena, C. Mainardi, S. Clemente, A. Varallo, S. Grasso, M. Buono, C. Oliviero, D. Manzi, M. Pugliese, A. Sarno. - In: IL NUOVO CIMENTO C. - ISSN 2037-4909. - 48:(2025), pp. 229.1-229.4. ( 110. National Congress of the Italian Physical Society : September, 9 - 13 Bologna 2024) [10.1393/ncc/i2025-25229-4].

3D image quality of a CBCT scanner for maxillofacial intrasurgical navigation

A. Sarno
Ultimo
2025

Abstract

This work aims to provide the first characterization of the Loop-X Cone-Beam CT scanner for maxillofacial applications, with a focus on patient dose and image quality in 3D acquisitions. Measurements were performed with adult and paediatric “head” protocols, using the dedicated Catphan604 phantom to evaluate the different parameters. Spatial resolution was assessed using the modular transfer function, with a limiting spatial frequency —evaluated at 10% of its maximum— of 0.60 mm−1 for the adult protocol and 0.53 mm−1 for the paediatric protocol (mean value). In addition, noise was characterized in terms of noise power spectrum (NPS) and contrast to noise ratio (CNR) was calculated for the adult protocol between simulated soft tissue and bone (23 ± 1) and between spinal disc and soft tissue (1 ± 1).
Settore PHYS-06/A - Fisica per le scienze della vita, l'ambiente e i beni culturali
2025
4-set-2025
Società Italiana di Fisica (SIF)
Article (author)
File in questo prodotto:
File Dimensione Formato  
ncc13393.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Licenza: Creative commons
Dimensione 292.93 kB
Formato Adobe PDF
292.93 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1220657
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? ND
  • OpenAlex ND
social impact