Background: Peripheral intravenous catheters (PIVCs) are the most commonly used invasive medical device, yet despite best efforts by end-users, PIVCs experience unacceptably high early failure rates. We aimed to design a new PIVC that reduces the early failure rate of in-dwelling PIVCs and we conducted preliminary tests to assess its efficacy and safety in a porcine model of intravenous access. Methods: We used computer-aided design and simulation to create a PIVC with a ramped tip geometry, which directs the infused fluid away from the vein wall; we called the design the FloRamp™. We created FloRamp prototypes (test device) and tested them against a market-leading device (BD Insyte™; control device) in a highly-controlled setting with five insertion sites per device in four pigs. We measured resistance to infusion and visual infusion phlebitis (VIP) every 6 h and terminated the experiment at 48 h. Veins were harvested for histology and seven pathological markers were assessed. Results: Computer simulations showed that the optimum FloRamp tip reduced maximum endothelial shear stress by 60%, from 12.7 Pa to 5.1 Pa, compared to a typical PIVC tip and improved the infusion dynamics of saline in the blood stream. In the animal study, we found that 2/5 of the control devices were occluded after 24 h, whereas all test devices remained patent and functional. The FloRamp created less resistance to infusion (0.73 ± 0.81 vs 0.47 ± 0.50, p = 0.06) and lower VIP scores (0.60 ± 0.93 vs 0.31 ± 0.70, p = 0.09) than the control device, although neither findings were significantly different. Histopathology revealed that 5/7 of the assessed markers were lower in veins with the FloRamp. Conclusions: Herein we report preliminary assessment of a novel PIVC design, which could be advantageous in clinical settings through decreased device occlusion and reduced early failure rates.

Design, development and preliminary assessment in a porcine model of a novel peripheral intravenous catheter aimed at reducing early failure rates / B.J. Doyle, L.J. Kelsey, C. Shelverton, G. Abbate, C. Ainola, N. Sato, S. Livingstone, M. Bouquet, M.R. Passmore, E.S. Wilson, S. Colombo, K. Sato, K. Liu, S. Heinsar, K. Wildi, P.J. Carr, J. Suen, J. Fraser, G. Li Bassi, S. Keogh. - In: JOURNAL OF VASCULAR ACCESS. - ISSN 1129-7298. - 25:3(2024 May), pp. 790-799. [10.1177/11297298221127760]

Design, development and preliminary assessment in a porcine model of a novel peripheral intravenous catheter aimed at reducing early failure rates

S. Colombo;G. Li Bassi;
2024

Abstract

Background: Peripheral intravenous catheters (PIVCs) are the most commonly used invasive medical device, yet despite best efforts by end-users, PIVCs experience unacceptably high early failure rates. We aimed to design a new PIVC that reduces the early failure rate of in-dwelling PIVCs and we conducted preliminary tests to assess its efficacy and safety in a porcine model of intravenous access. Methods: We used computer-aided design and simulation to create a PIVC with a ramped tip geometry, which directs the infused fluid away from the vein wall; we called the design the FloRamp™. We created FloRamp prototypes (test device) and tested them against a market-leading device (BD Insyte™; control device) in a highly-controlled setting with five insertion sites per device in four pigs. We measured resistance to infusion and visual infusion phlebitis (VIP) every 6 h and terminated the experiment at 48 h. Veins were harvested for histology and seven pathological markers were assessed. Results: Computer simulations showed that the optimum FloRamp tip reduced maximum endothelial shear stress by 60%, from 12.7 Pa to 5.1 Pa, compared to a typical PIVC tip and improved the infusion dynamics of saline in the blood stream. In the animal study, we found that 2/5 of the control devices were occluded after 24 h, whereas all test devices remained patent and functional. The FloRamp created less resistance to infusion (0.73 ± 0.81 vs 0.47 ± 0.50, p = 0.06) and lower VIP scores (0.60 ± 0.93 vs 0.31 ± 0.70, p = 0.09) than the control device, although neither findings were significantly different. Histopathology revealed that 5/7 of the assessed markers were lower in veins with the FloRamp. Conclusions: Herein we report preliminary assessment of a novel PIVC design, which could be advantageous in clinical settings through decreased device occlusion and reduced early failure rates.
biomaterials; Catheters; dialysis access; intensive care; new devices; nursing;
Settore MEDS-23/A - Anestesiologia
mag-2024
24-ott-2022
Article (author)
File in questo prodotto:
File Dimensione Formato  
doyle-et-al-2022-design-development-and-preliminary-assessment-in-a-porcine-model-of-a-novel-peripheral-intravenous.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Licenza: Nessuna licenza
Dimensione 2.12 MB
Formato Adobe PDF
2.12 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1189266
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 1
  • OpenAlex 1
social impact