Drusen are hallmarks of early and intermediate age-related macular degeneration (AMD) but their quantification remains a challenge. We compared automated drusen volume measurements between different OCT devices. We included 380 eyes from 200 individuals with bilateral intermediate (iAMD, n = 126), early (eAMD, n = 25) or no AMD (n = 49) from the MACUSTAR study. We assessed OCT scans from Cirrus (200 × 200 macular cube, 6 × 6 mm; Zeiss Meditec, CA) and Spectralis (20° × 20°, 25 B-scans; 30° × 25°, 241 B-scans; Heidelberg Engineering, Germany) devices. Sensitivity and specificity for drusen detection and differences between modalities were assessed with intra-class correlation coefficients (ICCs) and mean difference in a 5 mm diameter fovea-centered circle. Specificity was > 90% in the three modalities. In eAMD, we observed highest sensitivity in the denser Spectralis scan (68.1). The two different Spectralis modalities showed a significantly higher agreement in quantifying drusen volume in iAMD (ICC 0.993 [0.991–0.994]) than the dense Spectralis with Cirrus scan (ICC 0.807 [0.757–0.847]). Formulae for drusen volume conversion in iAMD between the two devices are provided. Automated drusen volume measures are not interchangeable between devices and softwares and need to be interpreted with the used imaging devices and software in mind. Accounting for systematic difference between methods increases comparability and conversion formulae are provided. Less dense scans did not affect drusen volume measurements in iAMD but decreased sensitivity for medium drusen in eAMD.

Comparability of automated drusen volume measurements in age-related macular degeneration: a MACUSTAR study report / D. Garzone, J.H. Terheyden, O. Morelle, M.W.M. Wintergerst, M. Saßmannshausen, S. Schmitz-Valckenberg, M. Pfau, S. Thiele, S. Poor, S. Leal, F.G. Holz, R.P. Finger, N. Null, H. Agostini, L. Altay, R. Atia, F. Bandello, P.G. Basile, C. Behning, M. Belmouhand, M. Berger, A. Binns, C.J.F. Boon, M. Böttger, C. Bouchet, J.E. Brazier, T. Butt, C. Carapezzi, J. Carlton, A. Carneiro, A. Charil, R. Coimbra, M. Cozzi, D.P. Crabb, J. Cunha-Vaz, C. Dahlke, L. De Sisternes, H. Dunbar, E. Fletcher, C. Francisco, M. Gutfleisch, R. Hogg, C.B. Hoyng, A. Kilani, J. Krätzschmar, L. Kühlewein, M. Larsen, Y.T.E. Lechanteur, U.F.O. Luhmann, A. Lüning, I. Marques, C. Martinho, G. Montesano, Z. Mulyukov, M. Paques, B. Parodi, M. Parravano, S. Penas, T. Peters, T. Peto, S. Priglinger, D. Rowen, G.S. Rubin, J. Sahel, C. Sánchez, O. Sander, M. Schmid, H. Schrinner-Fenske, J. Siedlecki, R. Silva, A. Skelly, E. Souied, G. Staurenghi, L. Stöhr, D.J. Taylor, A. Tufail, M. Varano, L. Vieweg, L. Wintergerst, A. Wolf, N. Zakaria. - In: SCIENTIFIC REPORTS. - ISSN 2045-2322. - 12:1(2022 Dec 19), pp. 21911.1-21911.10. [10.1038/s41598-022-26223-w]

Comparability of automated drusen volume measurements in age-related macular degeneration: a MACUSTAR study report

M. Cozzi;G. Montesano;G. Staurenghi;
2022

Abstract

Drusen are hallmarks of early and intermediate age-related macular degeneration (AMD) but their quantification remains a challenge. We compared automated drusen volume measurements between different OCT devices. We included 380 eyes from 200 individuals with bilateral intermediate (iAMD, n = 126), early (eAMD, n = 25) or no AMD (n = 49) from the MACUSTAR study. We assessed OCT scans from Cirrus (200 × 200 macular cube, 6 × 6 mm; Zeiss Meditec, CA) and Spectralis (20° × 20°, 25 B-scans; 30° × 25°, 241 B-scans; Heidelberg Engineering, Germany) devices. Sensitivity and specificity for drusen detection and differences between modalities were assessed with intra-class correlation coefficients (ICCs) and mean difference in a 5 mm diameter fovea-centered circle. Specificity was > 90% in the three modalities. In eAMD, we observed highest sensitivity in the denser Spectralis scan (68.1). The two different Spectralis modalities showed a significantly higher agreement in quantifying drusen volume in iAMD (ICC 0.993 [0.991–0.994]) than the dense Spectralis with Cirrus scan (ICC 0.807 [0.757–0.847]). Formulae for drusen volume conversion in iAMD between the two devices are provided. Automated drusen volume measures are not interchangeable between devices and softwares and need to be interpreted with the used imaging devices and software in mind. Accounting for systematic difference between methods increases comparability and conversion formulae are provided. Less dense scans did not affect drusen volume measurements in iAMD but decreased sensitivity for medium drusen in eAMD.
Settore MEDS-17/A - Malattie dell'apparato visivo
Settore MEDS-26/C - Scienze delle professioni sanitarie della riabilitazione
   Develop and validate appropriate and acceptable outcome measures in intermediate age-related macular degeneration for future interventional clinical trials - Sofia ref.: 116076
   MACUSTAR
   European Commission
   Horizon 2020 Framework Programme
   116076
19-dic-2022
Article (author)
File in questo prodotto:
File Dimensione Formato  
unpaywall-bitstream-1751454891.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Licenza: Creative commons
Dimensione 1.58 MB
Formato Adobe PDF
1.58 MB Adobe PDF Visualizza/Apri
s41598-023-30360-1.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Licenza: Creative commons
Dimensione 839.05 kB
Formato Adobe PDF
839.05 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1173376
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 9
  • ???jsp.display-item.citation.isi??? 10
  • OpenAlex 9
social impact