Purpose:To compare indocyanine green angiography and four different optical coherence tomography angiography (OCTA) devices and to test their reproducibility in the evaluation of choroidal neovascularization (CNV).MethodsThis study was an observational case series of Type 1 and Type 2 CNV presenting at the Eye Clinic, Department of Biomedical and Clinical Sciences, Luigi Sacco Hospital, University of Milan, imaged with indocyanine green angiography and four different OCTA devices: prototype PlexElite (Carl Zeiss Meditec), prototype Spectralis OCTA (Spectralis; Heidelberg Engineering, Heidelberg, Germany), Optovue RTVue XR Avanti (Optovue, Inc, Fremont, CA), and AngioPlex (Cirrus 5000 HD-OCT; Carl Zeiss Meditec, Inc, Dublin, CA). Optical coherence tomography angiography examinations were performed using a 3 × 3-mm or 6 × 6-mm volume scan pattern to capture the full lesion. The mean area of CNV, vessel density, and fractal dimension were compared between the devices.ResultsSeventeen eyes of 17 patients with CNV were enrolled. The CNV area was significantly larger in indocyanine green angiography than in the OCTA devices. Moreover, the four different instruments produced measurements with different mean values of CNV area, vessel density, and fractal dimension. The Bland-Altman analysis revealed that the limits of agreement for the comparisons were not acceptable.ConclusionThe choroidal neovascularization area is larger when imaged with standard indocyanine green angiography than with different OCTA devices with different algorithms, wavelengths, and scan patterns. The differences between several OCTA devices in the evaluation of quantitative parameters of CNV, such as CNV area, vessel density, and fractal dimension, suggest the necessity to standardize OCTA measurements for research and clinical practice.

Comparison between several optical coherence tomography angiography devices and indocyanine green angiography of choroidal neovascularization / F. Corvi, M. Cozzi, E. Barbolini, D. Nizza, M. Belotti, G. Staurenghi, A. Giani. - In: RETINA. - ISSN 0275-004X. - 40:5(2020 May), pp. 873-880. [10.1097/iae.0000000000002471]

Comparison between several optical coherence tomography angiography devices and indocyanine green angiography of choroidal neovascularization

M. Cozzi
Secondo
;
G. Staurenghi;A. Giani
Ultimo
2020

Abstract

Purpose:To compare indocyanine green angiography and four different optical coherence tomography angiography (OCTA) devices and to test their reproducibility in the evaluation of choroidal neovascularization (CNV).MethodsThis study was an observational case series of Type 1 and Type 2 CNV presenting at the Eye Clinic, Department of Biomedical and Clinical Sciences, Luigi Sacco Hospital, University of Milan, imaged with indocyanine green angiography and four different OCTA devices: prototype PlexElite (Carl Zeiss Meditec), prototype Spectralis OCTA (Spectralis; Heidelberg Engineering, Heidelberg, Germany), Optovue RTVue XR Avanti (Optovue, Inc, Fremont, CA), and AngioPlex (Cirrus 5000 HD-OCT; Carl Zeiss Meditec, Inc, Dublin, CA). Optical coherence tomography angiography examinations were performed using a 3 × 3-mm or 6 × 6-mm volume scan pattern to capture the full lesion. The mean area of CNV, vessel density, and fractal dimension were compared between the devices.ResultsSeventeen eyes of 17 patients with CNV were enrolled. The CNV area was significantly larger in indocyanine green angiography than in the OCTA devices. Moreover, the four different instruments produced measurements with different mean values of CNV area, vessel density, and fractal dimension. The Bland-Altman analysis revealed that the limits of agreement for the comparisons were not acceptable.ConclusionThe choroidal neovascularization area is larger when imaged with standard indocyanine green angiography than with different OCTA devices with different algorithms, wavelengths, and scan patterns. The differences between several OCTA devices in the evaluation of quantitative parameters of CNV, such as CNV area, vessel density, and fractal dimension, suggest the necessity to standardize OCTA measurements for research and clinical practice.
age-related macular degeneration; choroidal neovascularization; fluorescein angiography; fractal dimension; indocyanine green angiography; optical coherence tomography angiography; vessel density
Settore MEDS-17/A - Malattie dell'apparato visivo
Settore MEDS-26/C - Scienze delle professioni sanitarie della riabilitazione
mag-2020
Article (author)
File in questo prodotto:
File Dimensione Formato  
comparison_between_several_optical_coherence.11.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Licenza: Nessuna licenza
Dimensione 1.41 MB
Formato Adobe PDF
1.41 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1173350
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 42
  • ???jsp.display-item.citation.isi??? 40
  • OpenAlex 48
social impact