Aims Pulsed-field ablation (PFA) is a novel, myocardial-selective, non-thermal ablation modality used to target cardiac arrhythmias. Although prompt electrogram (EGM) signal disappearance is observed immediately after PFA application in the pulmonary veins, whether this finding results in adequate transmural lesions is unknown. The aim of this study is to check whether application repetition and catheter–tissue contact impact lesion formation during PFA. Methods A circular loop PFA catheter was used to deliver repeated energy applications with various levels of contact force. A bench- and results top vegetal potato model and a beating heart ventricular myocardial model were utilized to evaluate the impact of application repetition, contact force, and catheter repositioning on contiguity and lesion depth. Lesion development occurred over 18 h in the vegetal model and over 6 h in the porcine model. Lesion formation was found to be dependent on application repetition and contact. In porcine ventricles, single and multiple stacked applications led to a lesion depth of 3.5 ± 0.7 and 4.4 ± 1.3 mm, respectively (P = 0.002). Furthermore, the greater the catheter–tissue contact, the more contiguous and deeper the lesions in the vegetal model (1.0 ± 0.9 mm with no contact vs. 5.4 ± 1.4 mm with 30 g of force; P = 0.000Conclusion Pulsed-field ablation delivered via a circular catheter showed that both repetition and catheter contact led independently to deeper lesion formation. These findings indicate that endpoints for effective PFA are related more to PFA biophysics than to mere EGM attenuation.

Application repetition and electrode-tissue contact result in deeper lesions using a pulsed-field ablation circular variable loop catheter / L. Di Biase, J. Marazzato, T. Gomez, E. Byun, F. Zou, V. Grupposo, S. Mohanty, V.M. La Fazia, G. Ammirati, A. Lin, D.Y. Garcia, D. Della Rocca, A. Al Ahamad, M. Schiavone, A. Gasperetti, M. Freilich, J.C. Serna, G. Forleo, X. Liu, D. Lakkireddy, C. Tondo, A. Natale, X. Zhang. - In: EUROPACE. - ISSN 1532-2092. - 26:9(2024), pp. euae220.1-euae220.11. [10.1093/europace/euae220]

Application repetition and electrode-tissue contact result in deeper lesions using a pulsed-field ablation circular variable loop catheter

M. Schiavone;A. Gasperetti;C. Tondo;A. Natale;
2024

Abstract

Aims Pulsed-field ablation (PFA) is a novel, myocardial-selective, non-thermal ablation modality used to target cardiac arrhythmias. Although prompt electrogram (EGM) signal disappearance is observed immediately after PFA application in the pulmonary veins, whether this finding results in adequate transmural lesions is unknown. The aim of this study is to check whether application repetition and catheter–tissue contact impact lesion formation during PFA. Methods A circular loop PFA catheter was used to deliver repeated energy applications with various levels of contact force. A bench- and results top vegetal potato model and a beating heart ventricular myocardial model were utilized to evaluate the impact of application repetition, contact force, and catheter repositioning on contiguity and lesion depth. Lesion development occurred over 18 h in the vegetal model and over 6 h in the porcine model. Lesion formation was found to be dependent on application repetition and contact. In porcine ventricles, single and multiple stacked applications led to a lesion depth of 3.5 ± 0.7 and 4.4 ± 1.3 mm, respectively (P = 0.002). Furthermore, the greater the catheter–tissue contact, the more contiguous and deeper the lesions in the vegetal model (1.0 ± 0.9 mm with no contact vs. 5.4 ± 1.4 mm with 30 g of force; P = 0.000Conclusion Pulsed-field ablation delivered via a circular catheter showed that both repetition and catheter contact led independently to deeper lesion formation. These findings indicate that endpoints for effective PFA are related more to PFA biophysics than to mere EGM attenuation.
Catheter ablation; Irreversible electroporation; Pre-clinical model; Pulmonary vein isolation; Pulsed-field ablation
Settore MEDS-07/B - Malattie dell'apparato cardiovascolare
2024
Article (author)
File in questo prodotto:
File Dimensione Formato  
application.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 29.33 MB
Formato Adobe PDF
29.33 MB Adobe PDF Visualizza/Apri
application_compressed.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 505.61 kB
Formato Adobe PDF
505.61 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1158619
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 12
  • ???jsp.display-item.citation.isi??? 4
  • OpenAlex ND
social impact