Background: Catheter ablation using radiofrequency (RF) energy is an established treatment for ventricular tachycardia (VT). Tissue temperature is a key determinant of successful lesion creation, and yet, it is difficult to measure during conventional RF ablation because of the cooling effect of high-flow rate saline irrigation. The TRAC-VT study evaluated the safety and efficacy of a novel irrigated RF ablation system modulating power based on real-time tissue temperature. Methods: Patients with sustained monomorphic VT and structural heart disease (SHD) were enrolled. Catheter ablation was performed in temperature-control mode (irrigation 8 ml/min, temperature set-points 55 or 60 °C, and power output ≤ 50 W), with RF applications for ≤ 45 s. The primary safety endpoint was a composite of cardiovascular-specific serious procedure-related adverse events within 30 days post-ablation. The primary effectiveness endpoint was acute success (i.e., non-inducibility of all clinically relevant VTs). Results: Thirty-eight patients were enrolled with monomorphic VT (age 68 ± 12 years and 84% male), with an average of 1.7 ± 1.2 VTs targeted per patient. In total, 41 ± 23 RF applications per patient were delivered. Acute procedural success was 100% (95% CI, 91–100%). No primary safety endpoints were observed. Six-month follow-up was completed in 92% of patients with 81% (95% CI, 65–91%) freedom from sustained or treated VT. A repeat ablation was performed in three patients. Conclusions: Ablation of VT in SHD, using a temperature-controlled irrigated RF catheter, was safe and effective with a low rate of VT recurrence at 6 months.

Safety and efficacy of a temperature-controlled ablation system for ventricular tachycardia: Results from the TRAC-VT study / J. Kautzner, J. Moreno, C. Tondo, F. Anselme, J. Burrell, D. Becker, P. Peichl, I. Patchett, T. Dhanjal. - In: JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY. - ISSN 1572-8595. - 2025:(2025). [Epub ahead of print] [10.1007/s10840-025-01995-z]

Safety and efficacy of a temperature-controlled ablation system for ventricular tachycardia: Results from the TRAC-VT study

C. Tondo;
2025

Abstract

Background: Catheter ablation using radiofrequency (RF) energy is an established treatment for ventricular tachycardia (VT). Tissue temperature is a key determinant of successful lesion creation, and yet, it is difficult to measure during conventional RF ablation because of the cooling effect of high-flow rate saline irrigation. The TRAC-VT study evaluated the safety and efficacy of a novel irrigated RF ablation system modulating power based on real-time tissue temperature. Methods: Patients with sustained monomorphic VT and structural heart disease (SHD) were enrolled. Catheter ablation was performed in temperature-control mode (irrigation 8 ml/min, temperature set-points 55 or 60 °C, and power output ≤ 50 W), with RF applications for ≤ 45 s. The primary safety endpoint was a composite of cardiovascular-specific serious procedure-related adverse events within 30 days post-ablation. The primary effectiveness endpoint was acute success (i.e., non-inducibility of all clinically relevant VTs). Results: Thirty-eight patients were enrolled with monomorphic VT (age 68 ± 12 years and 84% male), with an average of 1.7 ± 1.2 VTs targeted per patient. In total, 41 ± 23 RF applications per patient were delivered. Acute procedural success was 100% (95% CI, 91–100%). No primary safety endpoints were observed. Six-month follow-up was completed in 92% of patients with 81% (95% CI, 65–91%) freedom from sustained or treated VT. A repeat ablation was performed in three patients. Conclusions: Ablation of VT in SHD, using a temperature-controlled irrigated RF catheter, was safe and effective with a low rate of VT recurrence at 6 months.
Cardiac arrhythmia; DiamondTemp; Radiofrequency catheter ablation; Temperature controlled; Ventricular tachycardia
Settore MEDS-07/B - Malattie dell'apparato cardiovascolare
2025
feb-2025
Article (author)
File in questo prodotto:
File Dimensione Formato  
Safety.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Licenza: Creative commons
Dimensione 665.91 kB
Formato Adobe PDF
665.91 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1147617
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
  • OpenAlex ND
social impact