Prompt Gamma Imaging (PGI) is a promising technique for range verification in Particle Therapy. This technique was already tested in clinical environment with a knife-edge-collimator camera for proton treatments but remains relatively unexplored for Carbon Ion Radiation Therapy (CIRT). Previous FLUKA simulations suggested that PG profile shifts could be detected in CIRT with a precision of 4 mm () for a particle statistic equal to C-ions using a 10 × 10 cm2 camera. An experimental campaign was carried out at CNAO (Pavia, Italy) to verify these results, using a knife-edge-collimator camera prototype based on a 5 × 5 cm2 pixelated LYSO crystal. PG profiles were measured irradiating a plastic phantom with a C-ion pencil beam at clinical energies and intensities, also moving the detector to extend the FOV to 13 × 5 cm2. The prototype detected Bragg-peak shifts with 4 mm precision for a statistic of C-ions (for the extended FOV), slightly larger than expected. Nevertheless, the detector demonstrated significant potential for verifying the precision in dose delivery following a treatment fraction, which remains fundamental in the clinical environment. For the first time to our knowledge, range verification based on PGI was applied to a C-ion beam at clinical energy and intensities.

First experimental verification of prompt gamma imaging with carbon ion irradiation / B. Idrissi Aicha, G. Borghi, A. Caracciolo, C. Riboldi, M. Carminati, M. Donetti, M. Pullia, S. Savazzi, F.E. Camera, C. Fiorini. - In: SCIENTIFIC REPORTS. - ISSN 2045-2322. - 14:1(2024 Oct 28), pp. 25750.1-25750.14. [10.1038/s41598-024-72870-6]

First experimental verification of prompt gamma imaging with carbon ion irradiation

F.E. Camera
Penultimo
;
2024

Abstract

Prompt Gamma Imaging (PGI) is a promising technique for range verification in Particle Therapy. This technique was already tested in clinical environment with a knife-edge-collimator camera for proton treatments but remains relatively unexplored for Carbon Ion Radiation Therapy (CIRT). Previous FLUKA simulations suggested that PG profile shifts could be detected in CIRT with a precision of 4 mm () for a particle statistic equal to C-ions using a 10 × 10 cm2 camera. An experimental campaign was carried out at CNAO (Pavia, Italy) to verify these results, using a knife-edge-collimator camera prototype based on a 5 × 5 cm2 pixelated LYSO crystal. PG profiles were measured irradiating a plastic phantom with a C-ion pencil beam at clinical energies and intensities, also moving the detector to extend the FOV to 13 × 5 cm2. The prototype detected Bragg-peak shifts with 4 mm precision for a statistic of C-ions (for the extended FOV), slightly larger than expected. Nevertheless, the detector demonstrated significant potential for verifying the precision in dose delivery following a treatment fraction, which remains fundamental in the clinical environment. For the first time to our knowledge, range verification based on PGI was applied to a C-ion beam at clinical energy and intensities.
Settore PHYS-01/A - Fisica sperimentale delle interazioni fondamentali e applicazioni
28-ott-2024
Article (author)
File in questo prodotto:
File Dimensione Formato  
s41598-024-72870-6.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 4.58 MB
Formato Adobe PDF
4.58 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1135196
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact