Purpose To evaluate the impact of temperature-controlled pars plana vitrectomy (PPV) on structural and functional outcomes in a rabbit eye model in vivo. Methods Ten healthy New Zealand White rabbits underwent temperature-controlled PPV in the right eye (group A), using a device specifically designed to heat the infusion fluid/air and integrated into the vitrectomy machine, and conventional PPV in the left eye (group B). Both eyes received ophthalmic examination and electroretinography (ERG) before and 1 week postoperatively. After 1-week ERG, rabbits were enucleated and then sacrificed. Histological and immunohistochemical examinations were performed on enucleated eyes and expression of glial fibrillary acidic protein (GFAP) and vimentin investigated. Results Postoperatively, only group B showed significantly decreased amplitude and increased latency of a-wave at 3 cds/m(2 )(p = 0.001 and 0.005, respectively). Significant increase of b-wave latency at 0.01 cds/m(2 )was detected in both groups (p = 0.019 and 0.023, respectively). Postoperatively, amplitude of oscillatory potentials (OPs) increased significantly in group A (p = 0.023) and decreased in group B. In both groups, OPs latency significantly increased at 1-week test (P < 0.05). A greater number of eyes without structural retinal alterations was detected in group A compared to group B (6 vs 5, respectively). GFAP expression was higher in group B than group A, even if the difference was not statistically significant. Conclusion Temperature-controlled PPV resulted in more favorable functional and structural outcomes in rabbit eyes compared with conventional PPV, supporting the potential beneficial role of the intraoperative management of intraocular temperature in vitreoretinal surgery.

Temperature control during pars plana vitrectomy / M.R. Romano, L. Barachetti, M. Ferrara, A. Mauro, L. Crepaldi, V. Bronzo, G. Franzo, G. Ravasio, C. Giudice. - In: GRAEFE'S ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY. - ISSN 0721-832X. - (2024 Sep 09). [Epub ahead of print] [10.1007/s00417-024-06631-6]

Temperature control during pars plana vitrectomy

V. Bronzo;G. Ravasio
Penultimo
;
C. Giudice
Ultimo
2024

Abstract

Purpose To evaluate the impact of temperature-controlled pars plana vitrectomy (PPV) on structural and functional outcomes in a rabbit eye model in vivo. Methods Ten healthy New Zealand White rabbits underwent temperature-controlled PPV in the right eye (group A), using a device specifically designed to heat the infusion fluid/air and integrated into the vitrectomy machine, and conventional PPV in the left eye (group B). Both eyes received ophthalmic examination and electroretinography (ERG) before and 1 week postoperatively. After 1-week ERG, rabbits were enucleated and then sacrificed. Histological and immunohistochemical examinations were performed on enucleated eyes and expression of glial fibrillary acidic protein (GFAP) and vimentin investigated. Results Postoperatively, only group B showed significantly decreased amplitude and increased latency of a-wave at 3 cds/m(2 )(p = 0.001 and 0.005, respectively). Significant increase of b-wave latency at 0.01 cds/m(2 )was detected in both groups (p = 0.019 and 0.023, respectively). Postoperatively, amplitude of oscillatory potentials (OPs) increased significantly in group A (p = 0.023) and decreased in group B. In both groups, OPs latency significantly increased at 1-week test (P < 0.05). A greater number of eyes without structural retinal alterations was detected in group A compared to group B (6 vs 5, respectively). GFAP expression was higher in group B than group A, even if the difference was not statistically significant. Conclusion Temperature-controlled PPV resulted in more favorable functional and structural outcomes in rabbit eyes compared with conventional PPV, supporting the potential beneficial role of the intraoperative management of intraocular temperature in vitreoretinal surgery.
heating device; hypothermia; intraocular temperature; pars plana vitrectomy; temperature-controlled pars plana vitrectomy; temperature fluctuations
Settore MVET-05/A - Clinica chirurgica veterinaria
Settore MVET-02/A - Patologia generale e anatomia patologica veterinaria
Settore MVET-03/A - Malattie infettive degli animali
9-set-2024
9-set-2024
Article (author)
File in questo prodotto:
File Dimensione Formato  
Temperature control during pars plana vitrectomy.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.37 MB
Formato Adobe PDF
1.37 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1102868
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 3
  • OpenAlex ND
social impact