Background: Variability remains a challenge in lupus anticoagulant (LA) testing. Objective: To validate LA test performance between Antiphospholipid Syndrome Alliance for Clinical Trials and International Networking (APS ACTION) Core laboratories and examine agreement in LA status between Core and local/hospital laboratories contributing patients to this prospective registry. Methods: Five Core laboratories used the same reagents, analyzer type, protocols, and characterized samples for LA validation. Non-anticoagulated registry samples were retested at the corresponding regional Core laboratories and anticoagulated samples at a single Core laboratory. Categorical agreement and discrepancies in LA status between Core and local/hospital laboratories were analyzed. Results: Clotting times for the reference/characterized plasmas used for normalized ratios were similar between Core laboratories (CV <4%); precision and agreement for LA positive/negative plasma were similar (all CV ≤5%) in the four laboratories that completed both parts of the validation exercise; 418 registry samples underwent LA testing. Agreement for LA positive/negative status between Core and local/hospital laboratories was observed in 87% (115/132) non-anticoagulated and 77% (183/237) anticoagulated samples. However, 28.7% (120/418) of samples showed discordance between the Core and local/hospital laboratories or equivocal LA results. Some of the results of the local/hospital laboratories might have been unreliable in 24.7% (41/166) and 23% (58/252) of the total non-anticoagulated and anticoagulated samples, respectively. Equivocal results by the Core laboratory might have also contributed to discordance. Conclusions: Laboratories can achieve good agreement in LA performance by use of the same reagents, analyzer type, and protocols. The standardized Core laboratory results underpin accurate interpretation of APS ACTION clinical data.

Comparison of real world and core laboratory lupus anticoagulant results from the Antiphospholipid Syndrome Alliance for Clinical Trials and International Networking (APS ACTION) clinical database and repository / M. Efthymiou, I.J. Mackie, P.J. Lane, D. Andrade, R. Willis, D. Erkan, S. Sciascia, S. Krillis, E. Bison, M. Borges Galhardo Vendramini, Z. Romay-Penabad, M. Qi, M. Tektonidou, A. Ugarte, C. Chighizola, H.M. Belmont, M.A. Aguirre, L. Ji, D.W. Branch, G. de Jesus, P.R. Fortin, L. Andreoli, M. Petri, R. Cervera, E. Rodriguez, J.S. Knight, T. Atsumi, J. Vega, E. Sevim, M.L. Bertolaccini, V. Pengo, H. Cohen. - In: JOURNAL OF THROMBOSIS AND HAEMOSTASIS. - ISSN 1538-7836. - 17:12(2019 Dec), pp. 2069-2080. [10.1111/jth.14596]

Comparison of real world and core laboratory lupus anticoagulant results from the Antiphospholipid Syndrome Alliance for Clinical Trials and International Networking (APS ACTION) clinical database and repository

C. Chighizola;
2019

Abstract

Background: Variability remains a challenge in lupus anticoagulant (LA) testing. Objective: To validate LA test performance between Antiphospholipid Syndrome Alliance for Clinical Trials and International Networking (APS ACTION) Core laboratories and examine agreement in LA status between Core and local/hospital laboratories contributing patients to this prospective registry. Methods: Five Core laboratories used the same reagents, analyzer type, protocols, and characterized samples for LA validation. Non-anticoagulated registry samples were retested at the corresponding regional Core laboratories and anticoagulated samples at a single Core laboratory. Categorical agreement and discrepancies in LA status between Core and local/hospital laboratories were analyzed. Results: Clotting times for the reference/characterized plasmas used for normalized ratios were similar between Core laboratories (CV <4%); precision and agreement for LA positive/negative plasma were similar (all CV ≤5%) in the four laboratories that completed both parts of the validation exercise; 418 registry samples underwent LA testing. Agreement for LA positive/negative status between Core and local/hospital laboratories was observed in 87% (115/132) non-anticoagulated and 77% (183/237) anticoagulated samples. However, 28.7% (120/418) of samples showed discordance between the Core and local/hospital laboratories or equivocal LA results. Some of the results of the local/hospital laboratories might have been unreliable in 24.7% (41/166) and 23% (58/252) of the total non-anticoagulated and anticoagulated samples, respectively. Equivocal results by the Core laboratory might have also contributed to discordance. Conclusions: Laboratories can achieve good agreement in LA performance by use of the same reagents, analyzer type, and protocols. The standardized Core laboratory results underpin accurate interpretation of APS ACTION clinical data.
antiphospholipid antibodies; antiphospholipid syndrome; lupus anticoagulant; Russell's viper venom time; thrombosis
Settore MED/18 - Chirurgia Generale
Settore MEDS-06/A - Chirurgia generale
dic-2019
Article (author)
File in questo prodotto:
File Dimensione Formato  
Efthymiou JTH 2019.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.03 MB
Formato Adobe PDF
1.03 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1086788
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 11
  • OpenAlex ND
social impact