We investigated the influence of model assumptions in GEANT4 Monte Carlo (MC) simulations for the calculation of monoenergetic and polyenergetic normalized glandular dose coefficients (DgN) in mammography, focussing on the effect of the skin thickness and composition, of the role of compression paddles and of the bremsstrahlung processes. We showed that selecting a skin thickness of 4 mm instead of 1.45 mm produced DgN values with deviations from 9% to 32% for x-ray spectra routinely adopted in mammography. Consideration of the bremsstrahlung radiation had a weak influence on monoenergetic DgN. Simulations (in the range 8-40 kVp) which included consideration of bremsstrahlung radiation, a skin thickness of 1.45 mm and a 2 mm thick compression paddles produced polyenergetic DgN coefficients up to 19% higher than corresponding literature data. Adding a 2 mm thick adipose layer between the skin layer and the radiosensitive portion of the breast produces polyenergetic DgN values up to 15% higher than those routinely adopted. These findings provide a quantitative estimate of the influence of model parameters on the calculation of the mean glandular dose in mammography.

A Monte Carlo study of monoenergetic and polyenergetic normalized glandular dose (DgN) coefficients in mammography / A. Sarno, G. Mettivier, F. Di Lillo, P. Russo. - In: PHYSICS IN MEDICINE AND BIOLOGY. - ISSN 0031-9155. - 62:1(2017), pp. 306-325. [10.1088/1361-6560/62/1/306]

A Monte Carlo study of monoenergetic and polyenergetic normalized glandular dose (DgN) coefficients in mammography

A. Sarno
Primo
;
2017

Abstract

We investigated the influence of model assumptions in GEANT4 Monte Carlo (MC) simulations for the calculation of monoenergetic and polyenergetic normalized glandular dose coefficients (DgN) in mammography, focussing on the effect of the skin thickness and composition, of the role of compression paddles and of the bremsstrahlung processes. We showed that selecting a skin thickness of 4 mm instead of 1.45 mm produced DgN values with deviations from 9% to 32% for x-ray spectra routinely adopted in mammography. Consideration of the bremsstrahlung radiation had a weak influence on monoenergetic DgN. Simulations (in the range 8-40 kVp) which included consideration of bremsstrahlung radiation, a skin thickness of 1.45 mm and a 2 mm thick compression paddles produced polyenergetic DgN coefficients up to 19% higher than corresponding literature data. Adding a 2 mm thick adipose layer between the skin layer and the radiosensitive portion of the breast produces polyenergetic DgN values up to 15% higher than those routinely adopted. These findings provide a quantitative estimate of the influence of model parameters on the calculation of the mean glandular dose in mammography.
GEANT4; mammography; mean glandular dose; Monte Carlo
Settore FIS/07 - Fisica Applicata(Beni Culturali, Ambientali, Biol.e Medicin)
2017
Article (author)
File in questo prodotto:
File Dimensione Formato  
Sarno_2017_Phys._Med._Biol._62_306.pdf

accesso riservato

Descrizione: Paper
Tipologia: Publisher's version/PDF
Dimensione 1.98 MB
Formato Adobe PDF
1.98 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1071595
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 34
  • ???jsp.display-item.citation.isi??? 33
  • OpenAlex ND
social impact