The estimation of the mean glandular dose in mammography using Monte Carlo simulations requires the calculation of the incident air kerma evaluated on the breast surface. In such a calculation, caution should be applied in considering explicitly the presence of the top compression paddle, since Compton scattering in this slab may produce a large spread of the incidence angles of x-ray photons on the scoring surface. Then, the calculation of the incident air kerma should contain the 'effective' area of the scoring surface, which takes into account the angle of incidence of photons on such a surface. Using Geant4 Monte Carlo simulations with a code previously validated according to the Task Group 195 of the American Association of Physicists in Medicine, we show that for typical x-ray spectra and energy range adopted in mammography, the resulting discrepancy in the calculation of the incident air kerma may lead to an overestimation from a minimum of 10% up to 12% of normalized dose coefficients and, hence, of the corresponding mean glandular dose if this contribution is not considered.
Air kerma calculation in Monte Carlo simulations for deriving normalized glandular dose coefficients in mammography / A. Sarno, G. Mettivier, P. Russo. - In: PHYSICS IN MEDICINE AND BIOLOGY. - ISSN 0031-9155. - 62:14(2017), pp. N337-N349. [10.1088/1361-6560/aa7016]
Air kerma calculation in Monte Carlo simulations for deriving normalized glandular dose coefficients in mammography
A. SarnoPrimo
;
2017
Abstract
The estimation of the mean glandular dose in mammography using Monte Carlo simulations requires the calculation of the incident air kerma evaluated on the breast surface. In such a calculation, caution should be applied in considering explicitly the presence of the top compression paddle, since Compton scattering in this slab may produce a large spread of the incidence angles of x-ray photons on the scoring surface. Then, the calculation of the incident air kerma should contain the 'effective' area of the scoring surface, which takes into account the angle of incidence of photons on such a surface. Using Geant4 Monte Carlo simulations with a code previously validated according to the Task Group 195 of the American Association of Physicists in Medicine, we show that for typical x-ray spectra and energy range adopted in mammography, the resulting discrepancy in the calculation of the incident air kerma may lead to an overestimation from a minimum of 10% up to 12% of normalized dose coefficients and, hence, of the corresponding mean glandular dose if this contribution is not considered.File | Dimensione | Formato | |
---|---|---|---|
Sarno_2017_Phys._Med._Biol._62_N337.pdf
accesso riservato
Descrizione: Article
Tipologia:
Publisher's version/PDF
Dimensione
658.81 kB
Formato
Adobe PDF
|
658.81 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.