We performed laboratory investigations of the image quality for 2-D [digital mammography (DM)] and 3-D [digital breast tomosynthesis (DBT) and computed tomography (CT)] in-line phase contrast imaging dedicated to the breast. We employed the cone-beam, microfocus CT scanner developed in-house. Planar imaging tests were carried out with BR50/50 and polymethylmethacrylate test objects, at tube voltages of 60 kV, 80 kV, and 100 kV. CT volume images, acquired at 80 kV at dose levels comparable to two-view mammography (mean glandular dose $\cong ~4$ mGy), as well as at 120 kV, were reconstructed from phase contrast projection images as well as from the corresponding phase retrieved projection images (Paganin's algorithm). We assumed 3-D spatial uniform distribution of the ratio $\delta /\beta $ of the X-ray optical coefficients of the sample material, at the effective energy of the X-ray beam. In 2-D imaging, all phantom details showed higher contrast after phase retrieval. In particular, contrast-to-noise ratio between a simulated mass and phantom background in phase retrieved images resulted up to 6.5 times higher than for absorption-based images. CT and DBT phase retrieval images of a homogeneous 50/50 test object at 120 kV produced higher signal-to-noise ratio of microcalcifications than in phase contrast DBT and CT of the same object, at the same glandular dose. In CT reconstructed volume images, microcalcifications down to 0.230-0.290-mm size presented an increased visibility after phase retrieval, in a two-components 50/50 breast phantom. For the first time, the advantages of in-line phase imaging in microfocus breast imaging have been shown with one setup for all techniques (DM, DBT, and breast CT). These results motivate further research on in-line phase contrast 2-D and 3-D imaging of the breast with a microfocus X-ray tube for potential application in the clinic of in-line phase contrast DM, DBT, and breast CT at dose levels comparable with corresponding attenuation-based exams.

In-Line Phase Contrast Mammography, Phase Contrast Digital Breast Tomosynthesis, and Phase Contrast Breast Computed Tomography with a Dedicated CT Scanner and a Microfocus X-Ray Tube: Experimental Phantom Study / G. Mettivier, F. Di Franco, A. Sarno, R. Castriconi, F. Di Lillo, K. Bliznakova, P. Russo. - In: IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES. - ISSN 2469-7311. - 5:6(2021), pp. 793-806. [10.1109/TRPMS.2020.3003380]

In-Line Phase Contrast Mammography, Phase Contrast Digital Breast Tomosynthesis, and Phase Contrast Breast Computed Tomography with a Dedicated CT Scanner and a Microfocus X-Ray Tube: Experimental Phantom Study

A. Sarno
;
R. Castriconi;
2021

Abstract

We performed laboratory investigations of the image quality for 2-D [digital mammography (DM)] and 3-D [digital breast tomosynthesis (DBT) and computed tomography (CT)] in-line phase contrast imaging dedicated to the breast. We employed the cone-beam, microfocus CT scanner developed in-house. Planar imaging tests were carried out with BR50/50 and polymethylmethacrylate test objects, at tube voltages of 60 kV, 80 kV, and 100 kV. CT volume images, acquired at 80 kV at dose levels comparable to two-view mammography (mean glandular dose $\cong ~4$ mGy), as well as at 120 kV, were reconstructed from phase contrast projection images as well as from the corresponding phase retrieved projection images (Paganin's algorithm). We assumed 3-D spatial uniform distribution of the ratio $\delta /\beta $ of the X-ray optical coefficients of the sample material, at the effective energy of the X-ray beam. In 2-D imaging, all phantom details showed higher contrast after phase retrieval. In particular, contrast-to-noise ratio between a simulated mass and phantom background in phase retrieved images resulted up to 6.5 times higher than for absorption-based images. CT and DBT phase retrieval images of a homogeneous 50/50 test object at 120 kV produced higher signal-to-noise ratio of microcalcifications than in phase contrast DBT and CT of the same object, at the same glandular dose. In CT reconstructed volume images, microcalcifications down to 0.230-0.290-mm size presented an increased visibility after phase retrieval, in a two-components 50/50 breast phantom. For the first time, the advantages of in-line phase imaging in microfocus breast imaging have been shown with one setup for all techniques (DM, DBT, and breast CT). These results motivate further research on in-line phase contrast 2-D and 3-D imaging of the breast with a microfocus X-ray tube for potential application in the clinic of in-line phase contrast DM, DBT, and breast CT at dose levels comparable with corresponding attenuation-based exams.
Breast computed tomography (CT); digital breast tomosynthesis (DBT); mammography; microfocus X-ray tube; phase contrast
Settore FIS/07 - Fisica Applicata(Beni Culturali, Ambientali, Biol.e Medicin)
   Three dimensional breast cancer models for X-ray Imaging research
   MaXIMA
   European Commission
   Horizon 2020 Framework Programme
   692097
2021
Article (author)
File in questo prodotto:
File Dimensione Formato  
In-Line_Phase_Contrast_Mammography_Phase_Contrast_Digital_Breast_Tomosynthesis_and_Phase_Contrast_Breast_Computed_Tomography_With_a_Dedicated_CT_Scanner_and_a_Microfocus_X-Ray_Tube_Experimental_P.pdf

accesso riservato

Descrizione: Article
Tipologia: Publisher's version/PDF
Dimensione 4.09 MB
Formato Adobe PDF
4.09 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1071575
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 0
social impact