We performed laboratory investigations of the image quality for 2-D [digital mammography (DM)] and 3-D [digital breast tomosynthesis (DBT) and computed tomography (CT)] in-line phase contrast imaging dedicated to the breast. We employed the cone-beam, microfocus CT scanner developed in-house. Planar imaging tests were carried out with BR50/50 and polymethylmethacrylate test objects, at tube voltages of 60 kV, 80 kV, and 100 kV. CT volume images, acquired at 80 kV at dose levels comparable to two-view mammography (mean glandular dose $\cong ~4$ mGy), as well as at 120 kV, were reconstructed from phase contrast projection images as well as from the corresponding phase retrieved projection images (Paganin's algorithm). We assumed 3-D spatial uniform distribution of the ratio $\delta /\beta $ of the X-ray optical coefficients of the sample material, at the effective energy of the X-ray beam. In 2-D imaging, all phantom details showed higher contrast after phase retrieval. In particular, contrast-to-noise ratio between a simulated mass and phantom background in phase retrieved images resulted up to 6.5 times higher than for absorption-based images. CT and DBT phase retrieval images of a homogeneous 50/50 test object at 120 kV produced higher signal-to-noise ratio of microcalcifications than in phase contrast DBT and CT of the same object, at the same glandular dose. In CT reconstructed volume images, microcalcifications down to 0.230-0.290-mm size presented an increased visibility after phase retrieval, in a two-components 50/50 breast phantom. For the first time, the advantages of in-line phase imaging in microfocus breast imaging have been shown with one setup for all techniques (DM, DBT, and breast CT). These results motivate further research on in-line phase contrast 2-D and 3-D imaging of the breast with a microfocus X-ray tube for potential application in the clinic of in-line phase contrast DM, DBT, and breast CT at dose levels comparable with corresponding attenuation-based exams.
In-Line Phase Contrast Mammography, Phase Contrast Digital Breast Tomosynthesis, and Phase Contrast Breast Computed Tomography with a Dedicated CT Scanner and a Microfocus X-Ray Tube: Experimental Phantom Study / G. Mettivier, F. Di Franco, A. Sarno, R. Castriconi, F. Di Lillo, K. Bliznakova, P. Russo. - In: IEEE TRANSACTIONS ON RADIATION AND PLASMA MEDICAL SCIENCES. - ISSN 2469-7311. - 5:6(2021), pp. 793-806. [10.1109/TRPMS.2020.3003380]
In-Line Phase Contrast Mammography, Phase Contrast Digital Breast Tomosynthesis, and Phase Contrast Breast Computed Tomography with a Dedicated CT Scanner and a Microfocus X-Ray Tube: Experimental Phantom Study
A. Sarno
;R. Castriconi;
2021
Abstract
We performed laboratory investigations of the image quality for 2-D [digital mammography (DM)] and 3-D [digital breast tomosynthesis (DBT) and computed tomography (CT)] in-line phase contrast imaging dedicated to the breast. We employed the cone-beam, microfocus CT scanner developed in-house. Planar imaging tests were carried out with BR50/50 and polymethylmethacrylate test objects, at tube voltages of 60 kV, 80 kV, and 100 kV. CT volume images, acquired at 80 kV at dose levels comparable to two-view mammography (mean glandular dose $\cong ~4$ mGy), as well as at 120 kV, were reconstructed from phase contrast projection images as well as from the corresponding phase retrieved projection images (Paganin's algorithm). We assumed 3-D spatial uniform distribution of the ratio $\delta /\beta $ of the X-ray optical coefficients of the sample material, at the effective energy of the X-ray beam. In 2-D imaging, all phantom details showed higher contrast after phase retrieval. In particular, contrast-to-noise ratio between a simulated mass and phantom background in phase retrieved images resulted up to 6.5 times higher than for absorption-based images. CT and DBT phase retrieval images of a homogeneous 50/50 test object at 120 kV produced higher signal-to-noise ratio of microcalcifications than in phase contrast DBT and CT of the same object, at the same glandular dose. In CT reconstructed volume images, microcalcifications down to 0.230-0.290-mm size presented an increased visibility after phase retrieval, in a two-components 50/50 breast phantom. For the first time, the advantages of in-line phase imaging in microfocus breast imaging have been shown with one setup for all techniques (DM, DBT, and breast CT). These results motivate further research on in-line phase contrast 2-D and 3-D imaging of the breast with a microfocus X-ray tube for potential application in the clinic of in-line phase contrast DM, DBT, and breast CT at dose levels comparable with corresponding attenuation-based exams.File | Dimensione | Formato | |
---|---|---|---|
In-Line_Phase_Contrast_Mammography_Phase_Contrast_Digital_Breast_Tomosynthesis_and_Phase_Contrast_Breast_Computed_Tomography_With_a_Dedicated_CT_Scanner_and_a_Microfocus_X-Ray_Tube_Experimental_P.pdf
accesso riservato
Descrizione: Article
Tipologia:
Publisher's version/PDF
Dimensione
4.09 MB
Formato
Adobe PDF
|
4.09 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.