We developed a scanner for micro computed tomography dedicated to the breast (BµCT) with a high resolution flat-panel detector and a microfocus X-ray tube. We evaluated the system spatial resolution via the 3D modulation transfer function (MTF). In addition to conventional absorption-based X-ray imaging, such a prototype showed capabilities for propagation-based phase-contrast and related edge enhancement effects in 3D imaging. The system limiting spatial resolution is 6.2 mm−1 (MTF at 10%) in the vertical direction and 3.8 mm−1 in the radial direction, values which compare favorably with the spatial resolution reached by mini focus breast CT scanners of other groups. The BµCT scanner was able to detect both microcalcification clusters and masses in an anthropomorphic breast phantom at a dose comparable to that of two-view mammography. The use of a breast holder is proposed in order to have 1–2 min long scan times without breast motion artifacts.
Cone-beam micro computed tomography dedicated to the breast / A. Sarno, G. Mettivier, F. Di Lillo, M. Cesarelli, P. Bifulco, P. Russo. - In: MEDICAL ENGINEERING & PHYSICS. - ISSN 1350-4533. - 38:12(2016), pp. 1449-1457. [10.1016/j.medengphy.2016.09.012]
Cone-beam micro computed tomography dedicated to the breast
A. SarnoPrimo
;
2016
Abstract
We developed a scanner for micro computed tomography dedicated to the breast (BµCT) with a high resolution flat-panel detector and a microfocus X-ray tube. We evaluated the system spatial resolution via the 3D modulation transfer function (MTF). In addition to conventional absorption-based X-ray imaging, such a prototype showed capabilities for propagation-based phase-contrast and related edge enhancement effects in 3D imaging. The system limiting spatial resolution is 6.2 mm−1 (MTF at 10%) in the vertical direction and 3.8 mm−1 in the radial direction, values which compare favorably with the spatial resolution reached by mini focus breast CT scanners of other groups. The BµCT scanner was able to detect both microcalcification clusters and masses in an anthropomorphic breast phantom at a dose comparable to that of two-view mammography. The use of a breast holder is proposed in order to have 1–2 min long scan times without breast motion artifacts.File | Dimensione | Formato | |
---|---|---|---|
1-s2.0-S1350453316302120-main.pdf
accesso riservato
Descrizione: Article
Tipologia:
Publisher's version/PDF
Dimensione
2.38 MB
Formato
Adobe PDF
|
2.38 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.