Background Screw misplacement incidence can be as high as 15-30% in spine deformity surgery, with possible devastating consequences. Some technical solutions to prevent misplacement require expensive devices. MySpine(TM) comprises a low-dose CT scan of the patient's spine to build a virtual model of the spine to plan the screw trajectories and a 3D-printed patient-specific guide system to prepare the screw trajectories and to implant the screws in the vertebrae in order to increase reproducibility and safety of the implants. The aim of this open-label, single-center, prospective randomized clinical trial with independent evaluation of outcomes was to compare the accuracy of free-hand insertion of pedicle screws to MySpine(TM) 3D-printed patient-specific guides. Methods Twenty-nine patients undergoing surgical correction for spinal deformity were randomized to Group A (pedicle screws implantation with MySpine(TM)) or Group B (free-hand implantation). Group A received 297 pedicle screws, and Group B 243 screws. Forty-three screws in Group A crossed over to free-hand implantation. Screw position was graded according to Gertzbein in grades 0, A, B or C, with grades 0 or A considered as safe area. Total fluoroscopy dose and time were compared in six patients of each group. Results Comparing the two study groups, we observed a statistically significant difference between the two groups (p<0.05), with 96.1% of screws in the safe area in Group A versus a 82.9% in Group B. Group-A patients had a mean effective dose of 0.23mSv compared to 0.82mSv in Group B. Patient-specific, 3D-printed pedicle screw guides increase safety in a wide spectrum of deformity conditions. In addition, the total radiation dose is reduced, even considering the need of a low-dose preoperative CT for surgical planning.

Pedicle screw insertion with patient-specific 3D-printed guides based on low-dose CT scan is more accurate than free-hand technique in spine deformity patients: a prospective, randomized clinical trial / R. Cecchinato, P. Berjano, A. Zerbi, M. Damilano, A. Redaelli, C. Lamartina. - In: EUROPEAN SPINE JOURNAL. - ISSN 0940-6719. - 28:7(2019 Jul), pp. 1712-1723. [10.1007/s00586-019-05978-3]

Pedicle screw insertion with patient-specific 3D-printed guides based on low-dose CT scan is more accurate than free-hand technique in spine deformity patients: a prospective, randomized clinical trial

R. Cecchinato
Primo
;
2019

Abstract

Background Screw misplacement incidence can be as high as 15-30% in spine deformity surgery, with possible devastating consequences. Some technical solutions to prevent misplacement require expensive devices. MySpine(TM) comprises a low-dose CT scan of the patient's spine to build a virtual model of the spine to plan the screw trajectories and a 3D-printed patient-specific guide system to prepare the screw trajectories and to implant the screws in the vertebrae in order to increase reproducibility and safety of the implants. The aim of this open-label, single-center, prospective randomized clinical trial with independent evaluation of outcomes was to compare the accuracy of free-hand insertion of pedicle screws to MySpine(TM) 3D-printed patient-specific guides. Methods Twenty-nine patients undergoing surgical correction for spinal deformity were randomized to Group A (pedicle screws implantation with MySpine(TM)) or Group B (free-hand implantation). Group A received 297 pedicle screws, and Group B 243 screws. Forty-three screws in Group A crossed over to free-hand implantation. Screw position was graded according to Gertzbein in grades 0, A, B or C, with grades 0 or A considered as safe area. Total fluoroscopy dose and time were compared in six patients of each group. Results Comparing the two study groups, we observed a statistically significant difference between the two groups (p<0.05), with 96.1% of screws in the safe area in Group A versus a 82.9% in Group B. Group-A patients had a mean effective dose of 0.23mSv compared to 0.82mSv in Group B. Patient-specific, 3D-printed pedicle screw guides increase safety in a wide spectrum of deformity conditions. In addition, the total radiation dose is reduced, even considering the need of a low-dose preoperative CT for surgical planning.
scoliosis; navigation; 3D printing; accuracy; spine deformity
Settore MED/33 - Malattie Apparato Locomotore
Settore MEDS-19/A - Malattie dell'apparato locomotore
lug-2019
Article (author)
File in questo prodotto:
File Dimensione Formato  
Cecchinato2019_Article_PedicleScrewInsertionWithPatie.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.33 MB
Formato Adobe PDF
1.33 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1061808
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 73
  • ???jsp.display-item.citation.isi??? 54
social impact