Background: This diagnostic prospective study compared the feasibility and diagnostic accuracy of Pocket-size Ultrasound Devices (PUDs) against standard ultrasound (US) in detecting liver steatosis using the controlled attenuation parameter (CAP) and liver biopsy as reference standards. Materials and Methods: Consecutive patients with chronic liver diseases were assessed for the presence of steatosis using PUD and US. A CAP cut-off value >275 dB/m was applied to establish ≥S1. A 26-patient subgroup underwent liver biopsy. PUD reproducibility was evaluated using Cohen's k statistic. Diagnostic accuracy of PUD and US was given as Sensibility (Sn), Specificity (Sp), Positive and Negative Predictive Values (PPV, NPV), positive and negative Likelihood Ratio (LR+, LR-). Results: 81 consecutive patients (69% males) with multiple etiologies were enroled. PUD inter-observer agreement was good (k 0.77, 95%CI 0.62–0.93). PUD and US identified ≥S1 according to CAP values respectively with Sn 0.87, Sp 0.61, PPV 0.49, NPV 0.91, LR+ 2.04, LR- 0.07, AUROC 0.74 and Sn 0.96, Sp 0.54, PPV 0.47, NPV 0.97, LR+ 2.10, LR- 0.07, AUROC 0.75. Conclusions: PUD shows good reproducibility and diagnostic accuracy in ruling liver steatosis out, representing a useful point-of-care tool to avail of hepatologists interested in excluding NAFLD, but with basic US skills.

Reproducibility and accuracy of a pocket-size ultrasound device in assessing liver steatosis / A. Costantino, A. Piagnani, R. Caccia, A. Sorge, M. Maggioni, R. Perbellini, F. Donato, R. D'Ambrosio, N.P. O Sed, L. Valenti, D. Prati, M. Vecchi, P. Lampertico, M. Fraquelli. - In: DIGESTIVE AND LIVER DISEASE. - ISSN 1590-8658. - (2023), pp. 1-7. [Epub ahead of print] [10.1016/j.dld.2023.11.014]

Reproducibility and accuracy of a pocket-size ultrasound device in assessing liver steatosis

A. Sorge;L. Valenti;M. Vecchi;P. Lampertico
Penultimo
;
2023

Abstract

Background: This diagnostic prospective study compared the feasibility and diagnostic accuracy of Pocket-size Ultrasound Devices (PUDs) against standard ultrasound (US) in detecting liver steatosis using the controlled attenuation parameter (CAP) and liver biopsy as reference standards. Materials and Methods: Consecutive patients with chronic liver diseases were assessed for the presence of steatosis using PUD and US. A CAP cut-off value >275 dB/m was applied to establish ≥S1. A 26-patient subgroup underwent liver biopsy. PUD reproducibility was evaluated using Cohen's k statistic. Diagnostic accuracy of PUD and US was given as Sensibility (Sn), Specificity (Sp), Positive and Negative Predictive Values (PPV, NPV), positive and negative Likelihood Ratio (LR+, LR-). Results: 81 consecutive patients (69% males) with multiple etiologies were enroled. PUD inter-observer agreement was good (k 0.77, 95%CI 0.62–0.93). PUD and US identified ≥S1 according to CAP values respectively with Sn 0.87, Sp 0.61, PPV 0.49, NPV 0.91, LR+ 2.04, LR- 0.07, AUROC 0.74 and Sn 0.96, Sp 0.54, PPV 0.47, NPV 0.97, LR+ 2.10, LR- 0.07, AUROC 0.75. Conclusions: PUD shows good reproducibility and diagnostic accuracy in ruling liver steatosis out, representing a useful point-of-care tool to avail of hepatologists interested in excluding NAFLD, but with basic US skills.
Controlled-attenuation parameter; Liver biopsy; Non-alcoholic fatty liver disease; Pocket size ultrasound devices; Steatosis; Ultrasound; Vibration-controlled transient elastography
Settore MED/09 - Medicina Interna
2023
27-nov-2023
Article (author)
File in questo prodotto:
File Dimensione Formato  
Reproducibility and accuracy of a pocket-size ultrasound device in.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 744.57 kB
Formato Adobe PDF
744.57 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1050411
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact