Background and aims: Prior reports have demonstrated a favourable safety and efficacy profile of the Micra leadless pacemaker over mid-term follow-up; however, long-term outcomes in real-world clinical practice remain unknown. Updated performance of the Micra VR leadless pacemaker through five years from the worldwide post-approval registry (PAR) was assessed. Methods: All Micra PAR patients undergoing implant attempts were included. Endpoints included system- or procedure-related major complications and system revision rate for any cause through 60 months post-implant. Rates were compared through 36 months post-implant to a reference dataset of 2667 transvenous pacemaker patients using Fine-Gray competing risk models. Results: 1809 patients were enrolled between July 2015 and March 2018 and underwent implant attempts from 179 centres in 23 countries with a median follow-up period of 51.1 months (IQR: 21.6-64.2). The major complication rate at 60 months was 4.5% [95% confidence interval (CI): 3.6%-5.5%] and was 4.1% at 36 months, which was significantly lower than the 8.5% rate observed for transvenous systems (HR: .47, 95% CI: .36-.61; P < .001). The all-cause system revision rate at 60 months was 4.9% (95% CI: 3.9%-6.1%). System revisions among Micra patients were mostly for device upgrades (41.2%) or elevated thresholds (30.6%). There were no Micra removals due to infection noted over the duration of follow-up. At 36 months, the system revision rate was significantly lower with Micra vs. transvenous systems (3.2% vs. 6.6%, P < .001). Conclusions: Long-term outcomes with the Micra leadless pacemaker continue to demonstrate low rates of major complications and system revisions and an extremely low incidence of infection.
Leadless pacemakers at 5-year follow-up: the Micra transcatheter pacing system post-approval registry / M.F. El-Chami, C. Garweg, N. Clementy, F. Al-Samadi, S. Iacopino, J.L. Martinez-Sande, P.R. Roberts, C. Tondo, J.B. Johansen, X. Vinolas-Prat, Y. Cha, E. Grubman, P. Bordachar, K. Stromberg, D.H. Fagan, J.P. Piccini. - In: EUROPEAN HEART JOURNAL. - ISSN 0195-668X. - (2024), pp. ehae101.1-ehae101.11. [10.1093/eurheartj/ehae101]
Leadless pacemakers at 5-year follow-up: the Micra transcatheter pacing system post-approval registry
C. Tondo;
2024
Abstract
Background and aims: Prior reports have demonstrated a favourable safety and efficacy profile of the Micra leadless pacemaker over mid-term follow-up; however, long-term outcomes in real-world clinical practice remain unknown. Updated performance of the Micra VR leadless pacemaker through five years from the worldwide post-approval registry (PAR) was assessed. Methods: All Micra PAR patients undergoing implant attempts were included. Endpoints included system- or procedure-related major complications and system revision rate for any cause through 60 months post-implant. Rates were compared through 36 months post-implant to a reference dataset of 2667 transvenous pacemaker patients using Fine-Gray competing risk models. Results: 1809 patients were enrolled between July 2015 and March 2018 and underwent implant attempts from 179 centres in 23 countries with a median follow-up period of 51.1 months (IQR: 21.6-64.2). The major complication rate at 60 months was 4.5% [95% confidence interval (CI): 3.6%-5.5%] and was 4.1% at 36 months, which was significantly lower than the 8.5% rate observed for transvenous systems (HR: .47, 95% CI: .36-.61; P < .001). The all-cause system revision rate at 60 months was 4.9% (95% CI: 3.9%-6.1%). System revisions among Micra patients were mostly for device upgrades (41.2%) or elevated thresholds (30.6%). There were no Micra removals due to infection noted over the duration of follow-up. At 36 months, the system revision rate was significantly lower with Micra vs. transvenous systems (3.2% vs. 6.6%, P < .001). Conclusions: Long-term outcomes with the Micra leadless pacemaker continue to demonstrate low rates of major complications and system revisions and an extremely low incidence of infection.File | Dimensione | Formato | |
---|---|---|---|
ehae101.pdf
accesso aperto
Tipologia:
Publisher's version/PDF
Dimensione
786.06 kB
Formato
Adobe PDF
|
786.06 kB | Adobe PDF | Visualizza/Apri |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.