Introduction: This multicenter prospective cohort study aimed to assess the safety and clinical and radiologic performance of the CLS® BreviusTM Stem with Kinectiv® Technology. Material and methods: A total of 222 consecutive subjects, recruited in five different centers, qualifying for primary total hip arthroplasty (THA), were enrolled in the study. All the subjects received the CLS® BreviusTM Stem with Kinectiv® Technology. All the enrolled study subjects underwent pre-operative clinical and radiographic evaluation. Additionally, all subjects underwent post-operative clinical, functional and radiographic evaluations at 6 months and 1, 2, 3, and 5 years. These evaluations included implant survival, pain and functional performance (Harris Hip Score [HHS], University of California, Los Angeles [UCLA] Activity Score, Oxford Hip Score), subject quality-of-life (EQ-5D), radiographic parameters, complications, and concentration of metal ions (aluminum and titanium) in blood. Results: No revisions were performed during the follow-up period. Of the 222 patients, only 76 completed the 5-year follow-up. Only 7 and 5 patients had aluminum and titanium 5-year evaluations, respectively. All the clinical parameters showed an overall improvement in the overtime measured with ANOVA for repeated measures; furthermore, the clinical scores showed a statistically significant improvement at 5 years with respect to pre-operative value (p < 0.001). Aluminum and titanium showed no variation for repeated measures at different time points (p > 0.05). A total of six complications were reported, of which only two were hip-related. Conclusions: The function of the CLS® BreviusTM Stem with Kinectiv® Technology indicated that subject well-being significantly increased following THA regardless of age, gender, BMI, previous surgery, primary diagnosis, and lifestyle.

Satisfactory clinical and radiologic outcomes with a new shorter and modular stem for end-stage hip osteoarthritis: an international prospective multicentre pilot study / R. D'Ambrosi, A. Toni, J. Czubak, J. Guadilla, L. Lieber, I. Mariani, N. Ursino. - In: SICOT-J. - ISSN 2426-8887. - 8:(2022 Aug), pp. 9.1-9.18. [10.1051/sicotj/2022005]

Satisfactory clinical and radiologic outcomes with a new shorter and modular stem for end-stage hip osteoarthritis: an international prospective multicentre pilot study

R. D'Ambrosi
Primo
;
2022

Abstract

Introduction: This multicenter prospective cohort study aimed to assess the safety and clinical and radiologic performance of the CLS® BreviusTM Stem with Kinectiv® Technology. Material and methods: A total of 222 consecutive subjects, recruited in five different centers, qualifying for primary total hip arthroplasty (THA), were enrolled in the study. All the subjects received the CLS® BreviusTM Stem with Kinectiv® Technology. All the enrolled study subjects underwent pre-operative clinical and radiographic evaluation. Additionally, all subjects underwent post-operative clinical, functional and radiographic evaluations at 6 months and 1, 2, 3, and 5 years. These evaluations included implant survival, pain and functional performance (Harris Hip Score [HHS], University of California, Los Angeles [UCLA] Activity Score, Oxford Hip Score), subject quality-of-life (EQ-5D), radiographic parameters, complications, and concentration of metal ions (aluminum and titanium) in blood. Results: No revisions were performed during the follow-up period. Of the 222 patients, only 76 completed the 5-year follow-up. Only 7 and 5 patients had aluminum and titanium 5-year evaluations, respectively. All the clinical parameters showed an overall improvement in the overtime measured with ANOVA for repeated measures; furthermore, the clinical scores showed a statistically significant improvement at 5 years with respect to pre-operative value (p < 0.001). Aluminum and titanium showed no variation for repeated measures at different time points (p > 0.05). A total of six complications were reported, of which only two were hip-related. Conclusions: The function of the CLS® BreviusTM Stem with Kinectiv® Technology indicated that subject well-being significantly increased following THA regardless of age, gender, BMI, previous surgery, primary diagnosis, and lifestyle.
hip arthroplasty; multicenter study; short stem; brevius CLS; blood ion levels
Settore MED/33 - Malattie Apparato Locomotore
ago-2022
Article (author)
File in questo prodotto:
File Dimensione Formato  
BREVIUS SICOT.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 282.78 kB
Formato Adobe PDF
282.78 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1022263
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
  • OpenAlex ND
social impact