A key issue for the reliable use of new devices for the laboratory control of oral anticoagulant therapy with the INR is their conformity to the calibration model. In the past, their adequacy has mostly been assessed empirically without reference to the calibration model and the use of International Reference Preparations (IRP) for thromboplastin. In this study we reviewed the requirements to be fulfilled and applied them to the calibration of a new near-patient testing device (TAS, Cardiovascular Diagnostics) which uses thromboplastin-containing test cards for determination of the INR. On each of 10 working days citrated whole blood and plasma samples were obtained from 2 healthy subjects and 6 patients on oral anticoagulants. PT testing on whole blood and plasma was done with the TAS and parallel testing for plasma by the manual technique with the IRP CRM 149S. Conformity to the calibration model was judged satisfactory if the following requirements were met: (i) there was a linear relationship between paired log-PTs (TAS vs CRM 149S); (ii) the regression line drawn through patients data points, passed through those of normals; (iii) the precision of the calibration expressed as the CV of the slope was < 3%. A good linear relationship was observed for calibration plots for plasma and whole blood (r = 0.98). Regression lines drawn through patients data points, passed through those of normals. The CVs of the slope were in both cases 2.2% and the ISIs were 0.965 and 1.000 for whole blood and plasma. In conclusion, our study shows that near-patient testing devices can be considered reliable tools to measure INR in patients on oral anticoagulants and provides guidelines for their evaluation.

Determination of the International Sensitivity Index of a new near-patient testing device to monitor oral anticoagulant therapy. Overview on the assessment of conformity to the calibration model / A. Tripodi, V. Chantarangkul, M. Clerici, B. Negri, P.M. Mannucci. - In: THROMBOSIS AND HAEMOSTASIS. - ISSN 0340-6245. - 78:2(1997), pp. 855-858.

Determination of the International Sensitivity Index of a new near-patient testing device to monitor oral anticoagulant therapy. Overview on the assessment of conformity to the calibration model

A. Tripodi;
1997

Abstract

A key issue for the reliable use of new devices for the laboratory control of oral anticoagulant therapy with the INR is their conformity to the calibration model. In the past, their adequacy has mostly been assessed empirically without reference to the calibration model and the use of International Reference Preparations (IRP) for thromboplastin. In this study we reviewed the requirements to be fulfilled and applied them to the calibration of a new near-patient testing device (TAS, Cardiovascular Diagnostics) which uses thromboplastin-containing test cards for determination of the INR. On each of 10 working days citrated whole blood and plasma samples were obtained from 2 healthy subjects and 6 patients on oral anticoagulants. PT testing on whole blood and plasma was done with the TAS and parallel testing for plasma by the manual technique with the IRP CRM 149S. Conformity to the calibration model was judged satisfactory if the following requirements were met: (i) there was a linear relationship between paired log-PTs (TAS vs CRM 149S); (ii) the regression line drawn through patients data points, passed through those of normals; (iii) the precision of the calibration expressed as the CV of the slope was < 3%. A good linear relationship was observed for calibration plots for plasma and whole blood (r = 0.98). Regression lines drawn through patients data points, passed through those of normals. The CVs of the slope were in both cases 2.2% and the ISIs were 0.965 and 1.000 for whole blood and plasma. In conclusion, our study shows that near-patient testing devices can be considered reliable tools to measure INR in patients on oral anticoagulants and provides guidelines for their evaluation.
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
Settore MED/09 - Medicina Interna
Settore MED/15 - Malattie del Sangue
1997
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/189558
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 40
  • ???jsp.display-item.citation.isi??? 34
social impact