Effects determining the energy and spatial resolution of a calorimeter based on liquid krypton have been studied. With cathode strips of 10 mm a spatial resolution of 0.4 mm has been obtained in a cosmic rays test. The energy resolution of the calorimeter (0.4 ton of krypton) has been measured with positrons, achieving a rms of 5.7% at E = 130 MeV and 1.7% at E = 1200 MeV. The measurements are compared to Monte Carlo simulations.
Investigation of an electromagnetic calorimeter based on liquid krypton / V. Aulchenko, S. Klimenko, G. Kolachev, L. Leontiev, A. Onuchin, V. Panin, Y. Pril, V. Rodyakin, A. Rylin, V. Tayursky, Y. Tikhonov, P. Cantoni, P. Frabetti, L. Stagni, G. Lo Bianco, F. Palombo, P. Manfredi, V. Re, V. Speziali. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 289:3(1990), pp. 468-474.
Investigation of an electromagnetic calorimeter based on liquid krypton
F. Palombo;
1990
Abstract
Effects determining the energy and spatial resolution of a calorimeter based on liquid krypton have been studied. With cathode strips of 10 mm a spatial resolution of 0.4 mm has been obtained in a cosmic rays test. The energy resolution of the calorimeter (0.4 ton of krypton) has been measured with positrons, achieving a rms of 5.7% at E = 130 MeV and 1.7% at E = 1200 MeV. The measurements are compared to Monte Carlo simulations.Pubblicazioni consigliate