We describe a new method for simultaneous determination of four hemoglobin derivatives (deoxyhemoglobin, Hb; oxyhemoglobin, HbO2; methemoglobin, Hb+; and carbon monoxide hemoglobin, HbCO) and total oxygen content in 10 microliters of whole blood. Percentage HbO2, HbCO, Hb+, total hemoglobin (Hbt), and oxygen capacity can also be obtained from the experimental data by simple calculations. Total analysis time is 1 min. Blood is diluted 100-fold with a buffer contained in a quasi-anaerobic cuvette, where simultaneous measurements of oxygen pressure (by a po2 electrode) and adsorbance (at 497, 565, and 620 nm) are made. The decrease of oxygen pressure, as recorded by the oxygen electrode, is proportional to the amount of deoxyhemoglobin. The concentrations of HbO2, HbCO, and Hb+ can be obtained from absorbance measurements at the specified wavelengths. The new method eliminates the use of short-path optical cells and, due to the low sample volume requirement, makes possible the automated measurement of hemoglobin derivatives and oxygen saturation in arterialized capillary blood.

Simultaneous determination of hemoglobin derivatives, oxygen content, oxygen capacity, and oxygen saturation in 10 microliters of whole blood / L. Rossi Bernardi, M. Perrella, M. Luzzana, M. Samaja, I. Raffaele. - In: CLINICAL CHEMISTRY. - ISSN 0009-9147. - 23:7(1977 Jul), pp. 1215-1225.

Simultaneous determination of hemoglobin derivatives, oxygen content, oxygen capacity, and oxygen saturation in 10 microliters of whole blood

M. Samaja
Penultimo
;
1977

Abstract

We describe a new method for simultaneous determination of four hemoglobin derivatives (deoxyhemoglobin, Hb; oxyhemoglobin, HbO2; methemoglobin, Hb+; and carbon monoxide hemoglobin, HbCO) and total oxygen content in 10 microliters of whole blood. Percentage HbO2, HbCO, Hb+, total hemoglobin (Hbt), and oxygen capacity can also be obtained from the experimental data by simple calculations. Total analysis time is 1 min. Blood is diluted 100-fold with a buffer contained in a quasi-anaerobic cuvette, where simultaneous measurements of oxygen pressure (by a po2 electrode) and adsorbance (at 497, 565, and 620 nm) are made. The decrease of oxygen pressure, as recorded by the oxygen electrode, is proportional to the amount of deoxyhemoglobin. The concentrations of HbO2, HbCO, and Hb+ can be obtained from absorbance measurements at the specified wavelengths. The new method eliminates the use of short-path optical cells and, due to the low sample volume requirement, makes possible the automated measurement of hemoglobin derivatives and oxygen saturation in arterialized capillary blood.
Settore BIO/10 - Biochimica
lug-1977
http://www.clinchem.org/content/23/7/1215.long
Article (author)
File in questo prodotto:
File Dimensione Formato  
03-77ClinChem_Hb_simultaneous_determination.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 5.93 MB
Formato Adobe PDF
5.93 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/175177
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 20
  • ???jsp.display-item.citation.isi??? ND
social impact