Background: Several closure devices are routinely used for percutaneous arterial access, while a relatively low number is available for the management of large bore venous accesses. The Woggle technique is a modification of the purse-string suture which was introduced several years ago in patients undergoing hemodialysis. Methods: A population of 45 patients who underwent transvenous femoral structural heart interventions was retrospectively evaluated. The Woggle technique consists of a purge string suture with a collar to maintain the tension as stable over time and a suture lock to tighten the suture. Results: Sheaths magnitude ranged from 8 French (F) to 14 F. A rapid post-procedural hemostasis was achieved in the whole population, and in 95% of cases, definite hemostasis was obtained after the first single release; the mean time of release was 302 ± 83 min. Although no relevant bleedings were reported, a significant reduction in hemoglobin levels was found in the whole population. This decrement was statistically significant only in the group with sheaths higher than 12 F. A single mild local hematoma was recorded in the group in which smaller sheaths were used. Seventy-two percent of patients were pre-treated with a dual antiplatelet therapy. Conclusions: The Woggle technique has shown to be a simple, effective, and safe approach for the management of large bore venous in percutaneous structural heart interventions.

The "Woggle" Technique for Venous Access Site Management: An Old Technique for a New Need / G. Tumminello, L. Barbieri, C. Avallone, N. Bellissimo, L. Mircoli, F. Colombo, M. Vicenzi, M. Ruscica, S. Carugo. - In: JOURNAL OF CLINICAL MEDICINE. - ISSN 2077-0383. - 12:18(2023 Sep 20), pp. 6087.1-6087.8. [10.3390/jcm12186087]

The "Woggle" Technique for Venous Access Site Management: An Old Technique for a New Need

M. Vicenzi
Writing – Review & Editing
;
M. Ruscica
Penultimo
Writing – Review & Editing
;
S. Carugo
Ultimo
Writing – Original Draft Preparation
2023

Abstract

Background: Several closure devices are routinely used for percutaneous arterial access, while a relatively low number is available for the management of large bore venous accesses. The Woggle technique is a modification of the purse-string suture which was introduced several years ago in patients undergoing hemodialysis. Methods: A population of 45 patients who underwent transvenous femoral structural heart interventions was retrospectively evaluated. The Woggle technique consists of a purge string suture with a collar to maintain the tension as stable over time and a suture lock to tighten the suture. Results: Sheaths magnitude ranged from 8 French (F) to 14 F. A rapid post-procedural hemostasis was achieved in the whole population, and in 95% of cases, definite hemostasis was obtained after the first single release; the mean time of release was 302 ± 83 min. Although no relevant bleedings were reported, a significant reduction in hemoglobin levels was found in the whole population. This decrement was statistically significant only in the group with sheaths higher than 12 F. A single mild local hematoma was recorded in the group in which smaller sheaths were used. Seventy-two percent of patients were pre-treated with a dual antiplatelet therapy. Conclusions: The Woggle technique has shown to be a simple, effective, and safe approach for the management of large bore venous in percutaneous structural heart interventions.
large bore venous access; percutaneous closure vascular systems; percutaneous structural heart intervention;
Settore MED/04 - Patologia Generale
Settore MED/11 - Malattie dell'Apparato Cardiovascolare
20-set-2023
Article (author)
File in questo prodotto:
File Dimensione Formato  
Gabriele Tumminello.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 623.73 kB
Formato Adobe PDF
623.73 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1004989
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 0
  • ???jsp.display-item.citation.isi??? 0
social impact