The Spes Low-energy Internal Conversion Electron Spectrometer (SLICES) is a new electron spectrometer designed to perform off-beam measurements at the beta decay station of the SPES facility at the Legnaro National Laboratories (LNL). SLICES consists of a large area Si(Li) detector cooled at -150 degrees C and a magnetic transport system. Four permanent N52 magnets, located around a central photon shield focus the electrons on the detector. In this paper, a detailed description of the SLICES components is presented. The strength of the measured magnetic field has been compared with calculations. The efficiency of the apparatus has been simulated and compared with the experimental data obtained using a radioactive Bi-207 source. The commissioning of the spectrometer at the CN accelerator at LNL is also presented.
SLICES : Spes Low-energy Internal Conversion Electron Spectrometer / N. Marchini, A. Nannini, M. Ottanelli, A. Saltarelli, M. Rocchini, G. Benzoni, E.R. Gamba, A. Goasduff, A. Gottardo, T. Krings, M. Perri. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 1020(2021), pp. 165860.1-165860.9. [10.1016/j.nima.2021.165860]
SLICES : Spes Low-energy Internal Conversion Electron Spectrometer
E.R. Gamba;
2021
Abstract
The Spes Low-energy Internal Conversion Electron Spectrometer (SLICES) is a new electron spectrometer designed to perform off-beam measurements at the beta decay station of the SPES facility at the Legnaro National Laboratories (LNL). SLICES consists of a large area Si(Li) detector cooled at -150 degrees C and a magnetic transport system. Four permanent N52 magnets, located around a central photon shield focus the electrons on the detector. In this paper, a detailed description of the SLICES components is presented. The strength of the measured magnetic field has been compared with calculations. The efficiency of the apparatus has been simulated and compared with the experimental data obtained using a radioactive Bi-207 source. The commissioning of the spectrometer at the CN accelerator at LNL is also presented.File | Dimensione | Formato | |
---|---|---|---|
1-s2.0-S0168900221008457-main.pdf
accesso riservato
Tipologia:
Publisher's version/PDF
Dimensione
1.42 MB
Formato
Adobe PDF
|
1.42 MB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.