Purpose: The options for haemostasis and dissection during modern thyroid surgery include the use of a bipolar vessel sealing system and ultrasonic technology. The aim of this study was to compare these energy-based devices (EBDs) with respect to their use in open thyroidectomy. Methods: The FOCUS Shear (F) and LigaSure LF1212 (L) were evaluated. A total of 182 consecutive patients scheduled for open thyroidectomy were prospectively randomised into two similarly sized groups. The parameters of operative time, morbidity, incision length, postoperative pain and hospital stay were analysed. Results: The main clinically primary endpoints of the two devices were identical in the study as the rate of nerve lesions, bleeding, drainage, operative time and postoperative calcium concentration with no significant differences with respect to the instrument utilised. The mean length of the incision was greater in the F group (p<0.05). Patients in the F group were more likely to complain of pain while swallowing (p<0.001). Early postoperative measurements of intact parathyroid hormone plasma levels revealed that although the levels were within the reference range, they were significantly lower in the F group (p<0.001). Oral calcium supplementation was significantly higher and more prolonged in the F group. Conclusions: The present study demonstrated no significant difference in the rates of postoperative morbidity associated with these two different EBDs used. Differences in clinically less significant were founded and focused on.

The safety of energy-based devices in open thyroidectomy : A prospective, randomised study comparing the LigaSure™ (LF1212) and the Harmonic® FOCUS / G. Dionigi, L. Boni, S. Rausei, F. Frattini, C.C. Ferrari, A. Mangano, A. Leotta, M. Franchin. - In: LANGENBECK'S ARCHIVES OF SURGERY. - ISSN 1435-2443. - 397:5(2012), pp. 817-823. [10.1007/s00423-011-0898-0]

The safety of energy-based devices in open thyroidectomy : A prospective, randomised study comparing the LigaSure™ (LF1212) and the Harmonic® FOCUS

G. Dionigi
;
L. Boni;
2012

Abstract

Purpose: The options for haemostasis and dissection during modern thyroid surgery include the use of a bipolar vessel sealing system and ultrasonic technology. The aim of this study was to compare these energy-based devices (EBDs) with respect to their use in open thyroidectomy. Methods: The FOCUS Shear (F) and LigaSure LF1212 (L) were evaluated. A total of 182 consecutive patients scheduled for open thyroidectomy were prospectively randomised into two similarly sized groups. The parameters of operative time, morbidity, incision length, postoperative pain and hospital stay were analysed. Results: The main clinically primary endpoints of the two devices were identical in the study as the rate of nerve lesions, bleeding, drainage, operative time and postoperative calcium concentration with no significant differences with respect to the instrument utilised. The mean length of the incision was greater in the F group (p<0.05). Patients in the F group were more likely to complain of pain while swallowing (p<0.001). Early postoperative measurements of intact parathyroid hormone plasma levels revealed that although the levels were within the reference range, they were significantly lower in the F group (p<0.001). Oral calcium supplementation was significantly higher and more prolonged in the F group. Conclusions: The present study demonstrated no significant difference in the rates of postoperative morbidity associated with these two different EBDs used. Differences in clinically less significant were founded and focused on.
Energy-based devices; Hypocalcaemia; Morbidity; Recurrent laryngeal nerve; Thyroidectomy; Adult; Aged; Blood Loss, Surgical; Equipment Design; Equipment Safety; Female; Follow-Up Studies; Hemostasis, Surgical; Humans; Ligation; Male; Middle Aged; Prospective Studies; Thyroid Neoplasms; Thyroidectomy; Treatment Outcome; Young Adult; Surgical Instruments
Settore MED/18 - Chirurgia Generale
2012
Article (author)
File in questo prodotto:
File Dimensione Formato  
Dionigi2012_Article_TheSafetyOfEnergy-basedDevices.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 156.5 kB
Formato Adobe PDF
156.5 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/882338
Citazioni
  • ???jsp.display-item.citation.pmc??? 16
  • Scopus 50
  • ???jsp.display-item.citation.isi??? 46
social impact