Background: Highly sensitive cardiac troponin assays (hs-cTn) are not available as point-of-care (POC) measurements. As rapid testing cannot be achieved at the expense of clinical performance, there is an urgent need to develop and rigorously validate POC hs-cTn. Konica Minolta (KM) has recently developed a surface plasmon-field enhanced fluorescence spectroscopy-based POC hs-cTn I system. Methods: We validated the analytical characteristics of the KM POC system according to the international guidelines. Results: Limit of blank (LoB) and limit of detection (LoD) were 0.35 and 0.62 ng/L, respectively, hs-cTn I concentrations corresponding to a total CV of 20%, 10% and 5% were 1.5, 3.9 and 11.0 ng/L, respectively. Method comparison studies showed that KM calibration was successfully traced to higher-order references. Limit of quantitation (LoQ), i.e. the hs-cTn I concentration having a total error of measurement of ≤34%, was 10.0 ng/L. The upper reference limit (URL) for 600 healthy blood donors was calculated at 12.2 ng/L (90% confidence interval [CI]: 9.2–39.2), while sex-partitioned URLs were 20.6 (males) and 10.7 ng/L (females), respectively (p < 0.0001). KM assay measured hs-cTn I concentrations >LoD in 65.7% of all reference individuals, in 76.7% of males and in 54.7% of females, respectively. Conclusions: The KM system joins the characteristics of POC systems to the analytical performance of hs-cTn.

Analytical validation of a highly sensitive point-of-care system for cardiac troponin I determination / F. Braga, E. Aloisio, A. Panzeri, T. Nakagawa, M. Panteghini. - In: CLINICAL CHEMISTRY AND LABORATORY MEDICINE. - ISSN 1434-6621. - 58:1(2019 Dec 18), pp. 138-145.

Analytical validation of a highly sensitive point-of-care system for cardiac troponin I determination

F. Braga
Primo
;
E. Aloisio
Secondo
;
M. Panteghini
Ultimo
2019

Abstract

Background: Highly sensitive cardiac troponin assays (hs-cTn) are not available as point-of-care (POC) measurements. As rapid testing cannot be achieved at the expense of clinical performance, there is an urgent need to develop and rigorously validate POC hs-cTn. Konica Minolta (KM) has recently developed a surface plasmon-field enhanced fluorescence spectroscopy-based POC hs-cTn I system. Methods: We validated the analytical characteristics of the KM POC system according to the international guidelines. Results: Limit of blank (LoB) and limit of detection (LoD) were 0.35 and 0.62 ng/L, respectively, hs-cTn I concentrations corresponding to a total CV of 20%, 10% and 5% were 1.5, 3.9 and 11.0 ng/L, respectively. Method comparison studies showed that KM calibration was successfully traced to higher-order references. Limit of quantitation (LoQ), i.e. the hs-cTn I concentration having a total error of measurement of ≤34%, was 10.0 ng/L. The upper reference limit (URL) for 600 healthy blood donors was calculated at 12.2 ng/L (90% confidence interval [CI]: 9.2–39.2), while sex-partitioned URLs were 20.6 (males) and 10.7 ng/L (females), respectively (p < 0.0001). KM assay measured hs-cTn I concentrations >LoD in 65.7% of all reference individuals, in 76.7% of males and in 54.7% of females, respectively. Conclusions: The KM system joins the characteristics of POC systems to the analytical performance of hs-cTn.
high sensitivity troponin; method evaluation; point of care testing
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
18-dic-2019
Article (author)
File in questo prodotto:
File Dimensione Formato  
[(CCLM)] Analytical validation of a highly sensitive point-of-care system for cardiac troponin I determination.pdf

Open Access dal 19/12/2020

Tipologia: Publisher's version/PDF
Dimensione 337.43 kB
Formato Adobe PDF
337.43 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/696698
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 14
  • ???jsp.display-item.citation.isi??? 12
social impact