Primary tissue failure of bioprosthetic mitral valves due to cusp perforations or ruptures is an unusual complication on short-term follow-up. An 88-year-old male with a known history of mitral regurgitation (MR) treated with bioprosthetic valve replacement in 2016 was referred to our center for recurrent heart failure. The two-dimensional (2D) transthoracic echocardiography documented an intraprosthetic jet of regurgitation without identifying a clear morphological mechanism, nor quantifying precisely the mitral insufficiency. 3D transesophageal echocardiography (TOE) with the tool FlexiSlice added relevant information by providing insights into the pathophysiological mechanisms of MR. The present case emphasizes the importance of 3D TOE as a fundamental tool for the diagnostic algorithm of bioprosthetic valves failure, even in the more demanding cases.

The Additional Value of Three-Dimensional Transesophageal Echocardiography in the Diagnosis of Unusual Complication of Bioprosthetic Mitral Valve / M. Evangelista, M. Barletta, A. Corciu, V. Mantovani, L. Paoli, M. Guazzi, M. Tusa. - In: JOURNAL OF CARDIOVASCULAR ECHOGRAPHY. - ISSN 2347-193X. - 29:2(2019 Jun), pp. 78-81. [10.4103/jcecho.jcecho_75_18]

The Additional Value of Three-Dimensional Transesophageal Echocardiography in the Diagnosis of Unusual Complication of Bioprosthetic Mitral Valve

M. Evangelista
Primo
;
M. Barletta
Secondo
;
V. Mantovani;M. Guazzi
Penultimo
;
2019

Abstract

Primary tissue failure of bioprosthetic mitral valves due to cusp perforations or ruptures is an unusual complication on short-term follow-up. An 88-year-old male with a known history of mitral regurgitation (MR) treated with bioprosthetic valve replacement in 2016 was referred to our center for recurrent heart failure. The two-dimensional (2D) transthoracic echocardiography documented an intraprosthetic jet of regurgitation without identifying a clear morphological mechanism, nor quantifying precisely the mitral insufficiency. 3D transesophageal echocardiography (TOE) with the tool FlexiSlice added relevant information by providing insights into the pathophysiological mechanisms of MR. The present case emphasizes the importance of 3D TOE as a fundamental tool for the diagnostic algorithm of bioprosthetic valves failure, even in the more demanding cases.
Bioprosthetic valve; flail; mitral regurgitation; three-dimensional transesophageal echocardiography
Settore MED/11 - Malattie dell'Apparato Cardiovascolare
giu-2019
Article (author)
File in questo prodotto:
File Dimensione Formato  
JCardiovascEchography29278-5554754_152547.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 853.76 kB
Formato Adobe PDF
853.76 kB Adobe PDF Visualizza/Apri