Purpose Our purpose is to analyze the outcomes of endovascular aneurysm repair (EVAR) in patients treated with Excluder endograft (W.L. Gore and Associate, Flagstaff, AZ), comparing second generation, featuring SIM-PULL delivery system (ExSP) and third generation, featuring C3 (ExC3), concerning intraprocedural data and long-term outcomes. Methods In our single-center, comparative study, we retrospectively analyzed all patients undergoing elective EVAR with Excluder from May 2008 to December 2015. This cohort was firstly divided according to the design of the endograft used, and then, two subgroups of complex procedures were identified according to International Standards. Preliminary end points were early- and mid-term outcomes. Primary end point was procedural data (i.e., procedural and fluoroscopy time, radiation dose (DAP), and contrast medium amount). Results The study included 64 patients (24 ExSP and 40 ExC3) with a mean follow-up of 31.6 ± 22.9 months. Patients in ExC3 group had significantly more risk factors (past or present history of smoking, P = 0.019), comorbidities (chronic heart failure and chronic kidney disease, both P = 0.032), as well as a more unfavorable anatomy (neck angulation, P = 0.035). Concerning preliminary outcome, no significant between-group difference was noted. As for intraoperative data, procedure duration was significantly shorter: 120 vs. 151 min (P = 0.002) in the overall population and 129 vs. 173 min (P = 0.004) in complex cases. A significant reduction was also found in fluoroscopy time and radiation exposure: 24,084 vs. 32,548 cGy/cm2 (P = 0.020) in the overall population and 26,770 vs. 41,104 cGy/cm2 (P = 0.003) in complex cases. No significant difference was found for contrast volume. Conclusions The study shows that new C3 excluder enables to reduce radiation exposure and procedural time compared to the previous device. C3 excluder results are comparable to those of the previous device in spite of more comorbidities and complex anatomy of the treated patients. Further studies are needed to assess device performance on longer-term follow-up.

Third- versus second-generation stent graft for endovascular aneurysm repair : a device-specific analysis / M. Bergonti, G. Teruzzi, G. Santagostino, L. Grancini, C. Ferrari, D. Trabattoni, A. Lualdi, A. Bartorelli. - In: ANNALS OF VASCULAR SURGERY. - ISSN 0890-5096. - 44(2017 Oct), pp. 67-76. ((Intervento presentato al 37. convegno National Congress of the Italian Society of Interventional Cardiology (GISE) tenutosi a Genova nel 2016 [10.1016/j.avsg.2017.03.176].

Third- versus second-generation stent graft for endovascular aneurysm repair : a device-specific analysis

M. Bergonti
;
G. Santagostino;A. Lualdi;A. Bartorelli
2017

Abstract

Purpose Our purpose is to analyze the outcomes of endovascular aneurysm repair (EVAR) in patients treated with Excluder endograft (W.L. Gore and Associate, Flagstaff, AZ), comparing second generation, featuring SIM-PULL delivery system (ExSP) and third generation, featuring C3 (ExC3), concerning intraprocedural data and long-term outcomes. Methods In our single-center, comparative study, we retrospectively analyzed all patients undergoing elective EVAR with Excluder from May 2008 to December 2015. This cohort was firstly divided according to the design of the endograft used, and then, two subgroups of complex procedures were identified according to International Standards. Preliminary end points were early- and mid-term outcomes. Primary end point was procedural data (i.e., procedural and fluoroscopy time, radiation dose (DAP), and contrast medium amount). Results The study included 64 patients (24 ExSP and 40 ExC3) with a mean follow-up of 31.6 ± 22.9 months. Patients in ExC3 group had significantly more risk factors (past or present history of smoking, P = 0.019), comorbidities (chronic heart failure and chronic kidney disease, both P = 0.032), as well as a more unfavorable anatomy (neck angulation, P = 0.035). Concerning preliminary outcome, no significant between-group difference was noted. As for intraoperative data, procedure duration was significantly shorter: 120 vs. 151 min (P = 0.002) in the overall population and 129 vs. 173 min (P = 0.004) in complex cases. A significant reduction was also found in fluoroscopy time and radiation exposure: 24,084 vs. 32,548 cGy/cm2 (P = 0.020) in the overall population and 26,770 vs. 41,104 cGy/cm2 (P = 0.003) in complex cases. No significant difference was found for contrast volume. Conclusions The study shows that new C3 excluder enables to reduce radiation exposure and procedural time compared to the previous device. C3 excluder results are comparable to those of the previous device in spite of more comorbidities and complex anatomy of the treated patients. Further studies are needed to assess device performance on longer-term follow-up.
Aged; Aged, 80 and over; Aortic Aneurysm, Abdominal; Aortography; Blood Vessel Prosthesis Implantation; Comorbidity; Contrast Media; Endovascular Procedures; Female; Humans; Italy; Male; Middle Aged; Operative Time; Prosthesis Design; Radiation Dosage; Radiation Exposure; Retrospective Studies; Risk Factors; Time Factors; Tomography, X-Ray Computed; Treatment Outcome; Blood Vessel Prosthesis; Stents; Surgery; Cardiology and Cardiovascular Medicine
Settore MED/11 - Malattie dell'Apparato Cardiovascolare
ott-2017
Article (author)
File in questo prodotto:
File Dimensione Formato  
Third_versus Second.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 533.56 kB
Formato Adobe PDF
533.56 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/633965
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact