We report the measurement of the longitudinal diffusion constant in liquid argon with the DarkSide-50 dual-phase time projection chamber. The measurement is performed at drift electric fields of 100 V/cm, 150 V/cm, and 200 V/cm using high statistics 39 Ar decays from atmospheric argon. We derive an expression to describe the pulse shape of the electroluminescence signal (S2) in dual-phase TPCs. The derived S2 pulse shape is fit to events from the uppermost portion of the TPC in order to characterize the radial dependence of the signal. The results are provided as inputs to the measurement of the longitudinal diffusion constant D L , which we find to be (4.12 ± 0.09) cm 2 /s for a selection of 140 keV electron recoil events in 200 V/cm drift field and 2.8 kV/cm extraction field. To study the systematics of our measurement we examine data sets of varying event energy, field strength, and detector volume yielding a weighted average value for the diffusion constant of (4.09 ± 0.12) cm 2 /s. The measured longitudinal diffusion constant is observed to have an energy dependence, and within the studied energy range the result is systematically lower than other results in the literature.

Electroluminescence pulse shape and electron diffusion in liquid argon measured in a dual-phase TPC / P. Agnes, I.F.M. Albuquerque, T. Alexander, A.K. Alton, D.M. Asner, M.P. Ave, H.O. Back, B. Baldin, G. Batignani, K. Biery, V. Bocci, G. Bonfini, W. Bonivento, M. Bossa, B. Bottino, F. Budano, S. Bussino, M. Cadeddu, M. Cadoni, F. Calaprice, A. Caminata, N. Canci, A. Candela, M. Caravati, M. Cariello, M. Carlini, M. Carpinelli, S. Catalanotti, V. Cataudella, P. Cavalcante, S. Cavuoti, A. Chepurnov, C. Cicalò, A.G. Cocco, G. Covone, D. D'Angelo, M. D'Incecco, D. D'Urso, S. Davini, A. De Candia, S. De Cecco, M. De Deo, G. De Filippis, G. De Rosa, M. De Vincenzi, P. Demontis, A.V. Derbin, A. Devoto, F. Di Eusanio, G. Di Pietro, C. Dionisi, E. Edkins, A. Fan, G. Fiorillo, K. Fomenko, D. Franco, F. Gabriele, A. Gabrieli, C. Galbiati, C. Ghiano, S. Giagu, C. Giganti, G.K. Giovanetti, A.M. Goretti, F. Granato, M. Gromov, M. Guan, Y. Guardincerri, M. Gulino, B.R. Hackett, K. Herner, D. Hughes, P. Humble, E.V. Hungerford, A. Ianni, I. James, T.N. Johnson, K. Keeter, C.L. Kendziora, I. Kochanek, G. Koh, D. Korablev, G. Korga, A. Kubankin, M. Kuss, X. Li, M. Lissia, B. Loer, G. Longo, Y. Ma, A.A. Machado, I.N. Machulin, A. Mandarano, S.M. Mari, J. Maricic, C.J. Martoff, A. Messina, P.D. Meyers, R. Milincic, A. Monte, M. Morrocchi, B.J. Mount, V.N. Muratova, P. Musico, A. Navrer Agasson, A. Nozdrina, A. Oleinik, M. Orsini, F. Ortica, L. Pagani, M. Pallavicini, L. Pandola, E. Pantic, E. Paoloni, F. Pazzona, K. Pelczar, N. Pelliccia, A. Pocar, S. Pordes, H. Qian, M. Razeti, A. Razeto, B. Reinhold, A.L. Renshaw, M. Rescigno, Q. Riffard, A. Romani, B. Rossi, N. Rossi, D. Sablone, O. Samoylov, W. Sands, S. Sanfilippo, M. Sant, C. Savarese, B. Schlitzer, E. Segreto, D.A. Semenov, A. Sheshukov, P.N. Singh, M.D. Skorokhvatov, O. Smirnov, A. Sotnikov, C. Stanford, G.B. Suffritti, Y. Suvorov, R. Tartaglia, G. Testera, A. Tonazzo, P. Trinchese, E.V. Unzhakov, M. Verducci, A. Vishneva, B. Vogelaar, M. Wada, T.J. Waldrop, S. Walker, H. Wang, Y. Wang, A.W. Watson, S. Westerdale, M.M. Wojcik, X. Xiang, X. Xiao, C. Yang, Z. Ye, C. Zhu, G. Zuzel. - In: NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT. - ISSN 0168-9002. - 904(2018 Oct 01), pp. 23-34. [10.1016/j.nima.2018.06.077]

Electroluminescence pulse shape and electron diffusion in liquid argon measured in a dual-phase TPC

D. D'Angelo;
2018

Abstract

We report the measurement of the longitudinal diffusion constant in liquid argon with the DarkSide-50 dual-phase time projection chamber. The measurement is performed at drift electric fields of 100 V/cm, 150 V/cm, and 200 V/cm using high statistics 39 Ar decays from atmospheric argon. We derive an expression to describe the pulse shape of the electroluminescence signal (S2) in dual-phase TPCs. The derived S2 pulse shape is fit to events from the uppermost portion of the TPC in order to characterize the radial dependence of the signal. The results are provided as inputs to the measurement of the longitudinal diffusion constant D L , which we find to be (4.12 ± 0.09) cm 2 /s for a selection of 140 keV electron recoil events in 200 V/cm drift field and 2.8 kV/cm extraction field. To study the systematics of our measurement we examine data sets of varying event energy, field strength, and detector volume yielding a weighted average value for the diffusion constant of (4.09 ± 0.12) cm 2 /s. The measured longitudinal diffusion constant is observed to have an energy dependence, and within the studied energy range the result is systematically lower than other results in the literature.
Electron diffusion constant; Liquid argon; Time projection chamber
Settore FIS/01 - Fisica Sperimentale
Settore FIS/04 - Fisica Nucleare e Subnucleare
1-ott-2018
Article (author)
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S0168900218308155-main.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.74 MB
Formato Adobe PDF
1.74 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
1802.01427.pdf

accesso aperto

Tipologia: Pre-print (manoscritto inviato all'editore)
Dimensione 2.56 MB
Formato Adobe PDF
2.56 MB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/631605
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 15
social impact