BACKGROUND Pauses in chest compressions (CCs) have a negative association with survival from cardiac arrest. Electrocardiographic (ECG) rhythm analysis and defibrillator charging are significant contributors to CC pauses. OBJECTIVE Accuracy of the Analysis During Compressions with Fast Reconfirmation (ADC-FR) algorithm, which features automated rhythm analysis and charging during CCs to reduce CC pauses, was retrospectively determined in a large database of ECGs from 2701 patients with out-of-hospital cardiac arrest. METHODS The ADC-FR algorithm generated a total of 7264 advisories, of which 3575 were randomly assigned to a development data set and 3689 to a test data set. With ADC-FR, a high-pass digital filter is used to remove CC artifacts, while the underlying ECG rhythm is automatically interpreted. When CCs are paused at the end of the 2-minute cardiopulmonary resuscitation interval, a 3-second reconfirmation analysis is performed using the artifact-free ECG to confirm the shock/no-shock advisory. The sensitivity and specificity of the ADC-FR algorithm in correctly identifying shockable/nonshockable rhythms during CCs were calculated. RESULTS In both data sets, the accuracy of the ADC-FR algorithm for each ECG rhythm exceeded the recommended performance goals, which apply to a standard artifact-free ECG analysis. Sensitivity and specificity were 97% and 99%, respectively, for the development data set and 95% and 99% for the test data set. CONCLUSION The ADC-FR algorithm is highly accurate in discriminating shockable and nonshockable rhythms and can be used to reduce CC pauses.

Cardiac rhythm analysis during ongoing cardiopulmonary resuscitation using the Analysis During Compressions with Fast reconfirmation technology / F. Fumagalli, A.E. Silver, Q. Tan, N. Zaidi, G. Ristagno. - In: HEART RHYTHM. - ISSN 1547-5271. - 15:2(2018 Feb), pp. 248-255. [10.1016/j.hrthm.2017.09.003]

Cardiac rhythm analysis during ongoing cardiopulmonary resuscitation using the Analysis During Compressions with Fast reconfirmation technology

G. Ristagno
2018

Abstract

BACKGROUND Pauses in chest compressions (CCs) have a negative association with survival from cardiac arrest. Electrocardiographic (ECG) rhythm analysis and defibrillator charging are significant contributors to CC pauses. OBJECTIVE Accuracy of the Analysis During Compressions with Fast Reconfirmation (ADC-FR) algorithm, which features automated rhythm analysis and charging during CCs to reduce CC pauses, was retrospectively determined in a large database of ECGs from 2701 patients with out-of-hospital cardiac arrest. METHODS The ADC-FR algorithm generated a total of 7264 advisories, of which 3575 were randomly assigned to a development data set and 3689 to a test data set. With ADC-FR, a high-pass digital filter is used to remove CC artifacts, while the underlying ECG rhythm is automatically interpreted. When CCs are paused at the end of the 2-minute cardiopulmonary resuscitation interval, a 3-second reconfirmation analysis is performed using the artifact-free ECG to confirm the shock/no-shock advisory. The sensitivity and specificity of the ADC-FR algorithm in correctly identifying shockable/nonshockable rhythms during CCs were calculated. RESULTS In both data sets, the accuracy of the ADC-FR algorithm for each ECG rhythm exceeded the recommended performance goals, which apply to a standard artifact-free ECG analysis. Sensitivity and specificity were 97% and 99%, respectively, for the development data set and 95% and 99% for the test data set. CONCLUSION The ADC-FR algorithm is highly accurate in discriminating shockable and nonshockable rhythms and can be used to reduce CC pauses.
Cardiac arrest; Rhythm analysis; Automated external defibrillation; Preshock pauses; Chest compression; Defibrillation
Settore MED/41 - Anestesiologia
feb-2018
Article (author)
File in questo prodotto:
File Dimensione Formato  
ADC-FR (1).pdf

accesso aperto

Tipologia: Post-print, accepted manuscript ecc. (versione accettata dall'editore)
Dimensione 731.44 kB
Formato Adobe PDF
731.44 kB Adobe PDF Visualizza/Apri
1-s2.0-S1547527117310822-main.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.36 MB
Formato Adobe PDF
1.36 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/620377
Citazioni
  • ???jsp.display-item.citation.pmc??? 7
  • Scopus 19
  • ???jsp.display-item.citation.isi??? 18
social impact