BACKGROUND: A major objective of the IFCC Committee on Education and Use of Biomarkers in Diabetes is to generate awareness and improvement of HbA1c assays through evaluation of the performance by countries and manufacturers. METHODS: Fresh whole blood and lyophilized hemolysate specimens manufactured from the same pool were used by 17 external quality assessment organizers to evaluate analytical performance of 2166 laboratories. Results were evaluated per country, per manufacturer, and per manufacturer and country combined according to criteria of the IFCC model for quality targets. RESULTS: At the country level with fresh whole blood specimens, 6 countries met the IFCC criterion, 2 did not, and 2 were borderline. With lyophilized hemolysates, 5 countries met the criterion, 2 did not, and 3 were borderline. At the manufacturer level using fresh whole blood specimens, 13 manufacturers met the criterion, 8 did not, and 3 were borderline. Using lyophilized hemolysates, 7 manufacturers met the criterion, 6 did not, and 3 were borderline. In both country and manufacturer groups, the major contribution to total error derived from between-laboratory variation. There were no substantial differences in performance between groups using fresh whole blood or lyophilized hemolysate samples. CONCLUSIONS: The state of the art is that 1 of 20 laboratories does not meet the IFCC criterion, but there are substantial differences between country and between manufacturer groups. Efforts to further improve quality should focus on reducing between-laboratory variation. With some limitations, fresh whole blood and well-defined lyophilized specimens are suitable for purpose.

EurA1c: The European HbA1c trial to investigate the performance of HbA1c assays in 2166 laboratories across 17 countries and 24 manufacturers by use of the IFCC model for quality targets / C. Weykamp, W. John, E. English, R. Erasmus, D. Sacks, C. Buchta, M. Mueller, Y. Lenga, M. Budina, J. Kratochvila, B. Friedecky, J. Siest, P. Kaiser, A. Haliassos, O. Panagiotakis, K. Makris, H. Graham, A. Kane, T. Smith, N. Barrett, L. Sciacovelli, M. Plebani, A. Faria, A. Cardoso, H. Correia, M. Alemany, C. Alsina, C. Gómez, G. Nordin, C. Persson, R. Fried, F. Akcadag, M. Akgöz, D. Aslan, S. Jones, A. Thomas, P. Gillery, S. Jaisson, A. Mosca, R. Paleari, R. Slingerland, J. Slootstra, S. Leppink, A. Elmgren, R. Little, S. Connolly, V. Makky, M. Nowicki, C. Siebelder, L. Schröer-Janssen, M. Te Winkel, I. de Graaf, E. Lenters-Westra. - In: CLINICAL CHEMISTRY. - ISSN 0009-9147. - 64:8(2018 Aug), pp. 1183-1192.

EurA1c: The European HbA1c trial to investigate the performance of HbA1c assays in 2166 laboratories across 17 countries and 24 manufacturers by use of the IFCC model for quality targets

M. Plebani
Membro del Collaboration Group
;
A. Mosca
Membro del Collaboration Group
;
R. Paleari
Membro del Collaboration Group
;
2018

Abstract

BACKGROUND: A major objective of the IFCC Committee on Education and Use of Biomarkers in Diabetes is to generate awareness and improvement of HbA1c assays through evaluation of the performance by countries and manufacturers. METHODS: Fresh whole blood and lyophilized hemolysate specimens manufactured from the same pool were used by 17 external quality assessment organizers to evaluate analytical performance of 2166 laboratories. Results were evaluated per country, per manufacturer, and per manufacturer and country combined according to criteria of the IFCC model for quality targets. RESULTS: At the country level with fresh whole blood specimens, 6 countries met the IFCC criterion, 2 did not, and 2 were borderline. With lyophilized hemolysates, 5 countries met the criterion, 2 did not, and 3 were borderline. At the manufacturer level using fresh whole blood specimens, 13 manufacturers met the criterion, 8 did not, and 3 were borderline. Using lyophilized hemolysates, 7 manufacturers met the criterion, 6 did not, and 3 were borderline. In both country and manufacturer groups, the major contribution to total error derived from between-laboratory variation. There were no substantial differences in performance between groups using fresh whole blood or lyophilized hemolysate samples. CONCLUSIONS: The state of the art is that 1 of 20 laboratories does not meet the IFCC criterion, but there are substantial differences between country and between manufacturer groups. Efforts to further improve quality should focus on reducing between-laboratory variation. With some limitations, fresh whole blood and well-defined lyophilized specimens are suitable for purpose.
clinical biochemistry; biochemistry (medical)
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
ago-2018
Article (author)
File in questo prodotto:
File Dimensione Formato  
2018 EurA1c trial CC.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.05 MB
Formato Adobe PDF
1.05 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/616449
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 47
  • ???jsp.display-item.citation.isi??? 33
social impact