Leakage of colonised oropharyngeal secretions across the tracheal tube cuff may cause iatrogenic pulmonary infection. We studied a novel 'add-on' cuff, which can be inserted over an existing tracheal tube and advanced into the subglottic region. The physical properties of the novel silicone cuff (BronchoGuard, Ciel Medical, USA) were evaluated in comparison with the Hi-Lo® tracheal tube. In a bench study, we identified saline inflation volumes required to transmit pressures between 15 and 30 cmH2 O against artificial tracheas of 18, 20 and 22 mm internal diameter. We computed cuff compliance, and minimal inflation volume to achieve air sealing during mechanical ventilation. Finally, we compared the leakage flow rate of artificial saliva across the novel cuff. On average, the mean (SD) inflation volumes necessary to transmit tracheal pressures of 15, 20, 25 and 30 cmH2 O were 4.1 (2.2), 4.4 (2.3), 4.6 (2.4) and 4.8 (2.4) ml for the novel cuff and 7.7 (2.5), 8.0 (2.6), 8.4 (2.6) and 8.7 (2.7) ml for the Hi-Lo tube, respectively (p < 0.001). The minimal inflation volumes to achieve air sealing were 3.8 (0.9) and 10.5 (2.1) ml (p < 0.001), which resulted in transmitted tracheal pressures of 8.3 (9.8) and 27.6 (34.8) cmH2 O (p < 0.001). Compliance was 0.026 (0.004) and 0.616 (0.324) ml.cmH2 0-1 , respectively (p < 0.001). Although massive leak was found when the novel cuff transmitted pressures ≤ 20 cmH2 O against the trachea, leakage was avoided with pressures ≥ 25 cmH2 O, owing to optimal contact between the cuff and the tracheal wall. In contrast, the standard cuff consistently leaked irrespective of the pressure. We conclude that the novel cuff has advantageous properties that warrant clinical corroboration.

In-vitro analysis of a novel 'add-on' silicone cuff to improve sealing properties of tracheal tubes / G. Li Bassi, C. Chiurazzi, E. Aguilera, C. Travierso, D. Battaglini, M. Yang, A. Motos, H. Yang, A. Meli, D. Marti, O.T. Ranzani, F. Blasi, P. Pelosi, D. Chiumello, A. Torres. - In: ANAESTHESIA. - ISSN 0003-2409. - 73:11(2018 Nov), pp. 1372-1381. [10.1111/anae.14413]

In-vitro analysis of a novel 'add-on' silicone cuff to improve sealing properties of tracheal tubes

C. Chiurazzi;F. Blasi;D. Chiumello
Penultimo
;
2018

Abstract

Leakage of colonised oropharyngeal secretions across the tracheal tube cuff may cause iatrogenic pulmonary infection. We studied a novel 'add-on' cuff, which can be inserted over an existing tracheal tube and advanced into the subglottic region. The physical properties of the novel silicone cuff (BronchoGuard, Ciel Medical, USA) were evaluated in comparison with the Hi-Lo® tracheal tube. In a bench study, we identified saline inflation volumes required to transmit pressures between 15 and 30 cmH2 O against artificial tracheas of 18, 20 and 22 mm internal diameter. We computed cuff compliance, and minimal inflation volume to achieve air sealing during mechanical ventilation. Finally, we compared the leakage flow rate of artificial saliva across the novel cuff. On average, the mean (SD) inflation volumes necessary to transmit tracheal pressures of 15, 20, 25 and 30 cmH2 O were 4.1 (2.2), 4.4 (2.3), 4.6 (2.4) and 4.8 (2.4) ml for the novel cuff and 7.7 (2.5), 8.0 (2.6), 8.4 (2.6) and 8.7 (2.7) ml for the Hi-Lo tube, respectively (p < 0.001). The minimal inflation volumes to achieve air sealing were 3.8 (0.9) and 10.5 (2.1) ml (p < 0.001), which resulted in transmitted tracheal pressures of 8.3 (9.8) and 27.6 (34.8) cmH2 O (p < 0.001). Compliance was 0.026 (0.004) and 0.616 (0.324) ml.cmH2 0-1 , respectively (p < 0.001). Although massive leak was found when the novel cuff transmitted pressures ≤ 20 cmH2 O against the trachea, leakage was avoided with pressures ≥ 25 cmH2 O, owing to optimal contact between the cuff and the tracheal wall. In contrast, the standard cuff consistently leaked irrespective of the pressure. We conclude that the novel cuff has advantageous properties that warrant clinical corroboration.
mechanical ventilation; polyvinylchloride; silicone; tracheal tube; ventilator-associated pneumonia
Settore MED/10 - Malattie dell'Apparato Respiratorio
nov-2018
Article (author)
File in questo prodotto:
File Dimensione Formato  
in vitro analysis of a novel add-on silicone cuff Bassi_et_al-2018-Anaesthesia.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 648.28 kB
Formato Adobe PDF
648.28 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/594525
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
social impact