The Capillarys HbA1c kit implemented on the Capillarys 2 Flex Piercing platform uses capillary electrophoresis to separate and quantify HbA1c in human blood. We performed an evaluation of this system by checking imprecision, relative bias and robustness respect to the analysis of samples with variable total hemoglobin concentrations. Intra-assay CVs were between 1.1% and 2.8% for HbA1c values between 35.8 mmol/mol and 96.4 mmol/mol. Inter-assay CVs, evaluated on control materials, were 2.0% and 2.4 % for high (68.6 mmol/mol) and low (36.8 mmol/mol) control levels, respectively. Results in three blood samples with various concentrations of HbA1c (36, 60 and 87 mmol/mol) were not affected by variation in total hemoglobin concentrations (between 40 to 180 g/L). Only at very low total hemoglobin concentration, the imprecision was slightly higher (CV 3.1%). The results obtained by capillary electrophoresis (y-method) were well correlated with those obtained by the HPLC Tosoh G8 (x-method) (y = 0.73 + 0.978x, r=0.998, n=100).

Valutazione del sistema Capillarys 2 Flex Piercing per la misura dell'emoglobina A1c = Evaluation of the Capillarys 2 Flex Piercing system for the determination of hemoglobin A1c (HbA1c) / R. Paleari, A. Mosca. - In: BIOCHIMICA CLINICA. - ISSN 0393-0564. - 38:2(2014), pp. 110-114.

Valutazione del sistema Capillarys 2 Flex Piercing per la misura dell'emoglobina A1c = Evaluation of the Capillarys 2 Flex Piercing system for the determination of hemoglobin A1c (HbA1c)

R. Paleari;A. Mosca
2014

Abstract

The Capillarys HbA1c kit implemented on the Capillarys 2 Flex Piercing platform uses capillary electrophoresis to separate and quantify HbA1c in human blood. We performed an evaluation of this system by checking imprecision, relative bias and robustness respect to the analysis of samples with variable total hemoglobin concentrations. Intra-assay CVs were between 1.1% and 2.8% for HbA1c values between 35.8 mmol/mol and 96.4 mmol/mol. Inter-assay CVs, evaluated on control materials, were 2.0% and 2.4 % for high (68.6 mmol/mol) and low (36.8 mmol/mol) control levels, respectively. Results in three blood samples with various concentrations of HbA1c (36, 60 and 87 mmol/mol) were not affected by variation in total hemoglobin concentrations (between 40 to 180 g/L). Only at very low total hemoglobin concentration, the imprecision was slightly higher (CV 3.1%). The results obtained by capillary electrophoresis (y-method) were well correlated with those obtained by the HPLC Tosoh G8 (x-method) (y = 0.73 + 0.978x, r=0.998, n=100).
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
2014
Article (author)
File in questo prodotto:
File Dimensione Formato  
B_25_Sebia Capillarys.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 445.3 kB
Formato Adobe PDF
445.3 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/511915
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 0
  • ???jsp.display-item.citation.isi??? ND
social impact