Purpose: We investigated the effects of periodical high pressure breaths (SIGH) or biphasic positive pressure ventilation (BIPAP) during helmet continuous positive airway pressure (CPAP) in patients with acute hypoxic respiratory failure. Methods: We used a recently developed electromechanical expiratory valve (TwinPAP®, StarMed, Mirandola, Italy), which is time-cycled between two customizable positive end-expiratory pressure (PEEP) levels. We studied 21 patients (67 ± 17 years old) undergoing helmet CPAP. Continuous flow CPAP system was set at 60 l/min flow rate while maintaining clinical FiO2 (51 ± 15%). Five steps, lasting 1 h each, were applied: (1) spontaneous breathing with PEEP 0 cmH2O (SB), (2) CPAP with PEEP 8 cmH2O (CPAPbasal), (3) low PEEP, 8 cmH2O, for 25 s and high PEEP, 25 cmH2O, for 5 s (SIGH), (4) low PEEP, 8 cmH2O, for 3 s and high PEEP, 20 cmH2O, for 3 s (BIPAP), (5) CPAP with PEEP 8 cmH2O (CPAPfinal). We randomized the sequence of SIGH and BIPAP. Results: PaO2 was significantly higher during all steps compared to SB. When compared to CPAP basal, both SIGH and BIPAP induced a further increase in PaO 2. PaO2 during SIGH and BIPAP were not different. The oxygenation improvement was maintained during CPAPfinal. Conclusions: Superimposed, nonsynchronized positive pressure breaths delivered during helmet CPAP by means of the TwinPAP® system may improve oxygenation in patients with acute hypoxemic respiratory failure, even at a rate as low as two breaths per minute.
Continuous flow biphasic positive airway pressure by helmet in patients with acute hypoxic respiratory failure : effect on oxygenation / S. Isgrò, A. Zanella, C. Sala, G. Grasselli, G. Foti, A. Pesenti, N. Patroniti. - In: INTENSIVE CARE MEDICINE. - ISSN 0342-4642. - 36:10(2010), pp. 1688-1694.
Continuous flow biphasic positive airway pressure by helmet in patients with acute hypoxic respiratory failure : effect on oxygenation
A. Zanella;G. Grasselli;A. PesentiPenultimo
;
2010
Abstract
Purpose: We investigated the effects of periodical high pressure breaths (SIGH) or biphasic positive pressure ventilation (BIPAP) during helmet continuous positive airway pressure (CPAP) in patients with acute hypoxic respiratory failure. Methods: We used a recently developed electromechanical expiratory valve (TwinPAP®, StarMed, Mirandola, Italy), which is time-cycled between two customizable positive end-expiratory pressure (PEEP) levels. We studied 21 patients (67 ± 17 years old) undergoing helmet CPAP. Continuous flow CPAP system was set at 60 l/min flow rate while maintaining clinical FiO2 (51 ± 15%). Five steps, lasting 1 h each, were applied: (1) spontaneous breathing with PEEP 0 cmH2O (SB), (2) CPAP with PEEP 8 cmH2O (CPAPbasal), (3) low PEEP, 8 cmH2O, for 25 s and high PEEP, 25 cmH2O, for 5 s (SIGH), (4) low PEEP, 8 cmH2O, for 3 s and high PEEP, 20 cmH2O, for 3 s (BIPAP), (5) CPAP with PEEP 8 cmH2O (CPAPfinal). We randomized the sequence of SIGH and BIPAP. Results: PaO2 was significantly higher during all steps compared to SB. When compared to CPAP basal, both SIGH and BIPAP induced a further increase in PaO 2. PaO2 during SIGH and BIPAP were not different. The oxygenation improvement was maintained during CPAPfinal. Conclusions: Superimposed, nonsynchronized positive pressure breaths delivered during helmet CPAP by means of the TwinPAP® system may improve oxygenation in patients with acute hypoxemic respiratory failure, even at a rate as low as two breaths per minute.File | Dimensione | Formato | |
---|---|---|---|
220_icm_2010_3.pdf
accesso riservato
Tipologia:
Publisher's version/PDF
Dimensione
374.8 kB
Formato
Adobe PDF
|
374.8 kB | Adobe PDF | Visualizza/Apri Richiedi una copia |
Pubblicazioni consigliate
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.