Background: This investigation used image data generated by an anthropomorphic phantom with a cardiac insert for a comparison between two solid state cameras: D-SPECT and D530c. Methods: For each camera, two sets (with and without a simulated transmural defect (TD)) of scans were acquired starting from the in vivo standard count statistics in the left ventricle (LV). Other two acquisitions corresponding to 150% and 50% of the reference count statistics were acquired. Five performance indices related to spatial resolution, contrast, and contrast-to-noise ratio (CNR) were analyzed. Results: D-SPECT showed a lower LV wall thickness and an inferior sharpness than D530c. No significant differences were found in terms of contrast between LV wall and the inner cavity, TD contrast or CNR. No significant differences were observed in CNR when moving from the reference level of count statistics down to 50% or up to 150% of the counts acquired on the LV. Conclusions: Our results show that D-SPECT and D530c have different performances. The lack of differences in the image performance indices along the range of count statistics explored, indicates that there is the possibility for a further reduction in the injected activity and/or the acquisition time, for both systems.

Comparative analysis of cadmium-zincum-telluride cameras dedicated to myocardial perfusion SPECT : a phantom study / O. Zoccarato, D. Lizio, A. Savi, L. Indovina, C. Scabbio, L. Leva, A. Del Sole, C. Marcassa, R. Matheoud, M. Lecchi, M. Brambilla. - In: JOURNAL OF NUCLEAR CARDIOLOGY. - ISSN 1071-3581. - 23:4(2016 Aug), pp. 885-893. [10.1007/s12350-015-0203-7]

Comparative analysis of cadmium-zincum-telluride cameras dedicated to myocardial perfusion SPECT : a phantom study

C. Scabbio;A. Del Sole;M. Lecchi
Penultimo
;
2016

Abstract

Background: This investigation used image data generated by an anthropomorphic phantom with a cardiac insert for a comparison between two solid state cameras: D-SPECT and D530c. Methods: For each camera, two sets (with and without a simulated transmural defect (TD)) of scans were acquired starting from the in vivo standard count statistics in the left ventricle (LV). Other two acquisitions corresponding to 150% and 50% of the reference count statistics were acquired. Five performance indices related to spatial resolution, contrast, and contrast-to-noise ratio (CNR) were analyzed. Results: D-SPECT showed a lower LV wall thickness and an inferior sharpness than D530c. No significant differences were found in terms of contrast between LV wall and the inner cavity, TD contrast or CNR. No significant differences were observed in CNR when moving from the reference level of count statistics down to 50% or up to 150% of the counts acquired on the LV. Conclusions: Our results show that D-SPECT and D530c have different performances. The lack of differences in the image performance indices along the range of count statistics explored, indicates that there is the possibility for a further reduction in the injected activity and/or the acquisition time, for both systems.
cadmium-zincum-telluride; cardiac; SPECT; cardiology and cardiovascular medicine; radiology, nuclear medicine and imaging
Settore MED/36 - Diagnostica per Immagini e Radioterapia
Settore FIS/07 - Fisica Applicata(Beni Culturali, Ambientali, Biol.e Medicin)
ago-2016
2-lug-2015
Article (author)
File in questo prodotto:
File Dimensione Formato  
Zoccarato et al. - 2015 - Comparative analysis of cadmium-zincum-telluride cameras dedicated to myocardial perfusion SPECT A phantom stu.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.8 MB
Formato Adobe PDF
1.8 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/292738
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 24
  • ???jsp.display-item.citation.isi??? 17
social impact