This work deals with the development of a new gamma detector based on Silicon Drift Detectors (SDDs) to readout large LaBr3:Ce scintillators for gamma-ray spectroscopy and imaging applications. The research is supported by the European Space Agency through the Technology Research Programme (TRP) and by Istituto Nazionale di Fisica Nucleare (INFN) within the Gamma project. The SDDs, produced at Fondazione Bruno Kessler (FBK) semiconductor laboratories, are designed as monolithic arrays of 3 × 3 units, each one of an active area of 8 mm × 8 mm (overall area of 26 mm × 26 mm). The readout electronics and the architecture of the camera are briefly described and then first experimental results coupling the SDD array with a 1'' × 1'' LaBr3:Ce scintillator are reported. An energy resolution of 3% FWHM at 662 keV has been measured at -20°C, better than coupling the same scintillator with a photomultiplier tube. The same scintillator is also used to evaluate position sensitivity with a 1 mm collimated Cs-137 source. The main difficulty in determining the position of the gamma-ray interaction in the crystal is associated to the high thickness/diameter ratio of the crystal (1:1) and the use of reflectors on all lateral and top sides the crystal. This last choice enhances energy resolution but makes imaging capability more challenging because light is spread over all photodetectors. Preliminary results show that the camera is able to detect shifts in the measured signals, when the source is moved with steps of 5 mm. A modified version of the centroid method is finally implemented to evaluate the imaging capability of the system.

Development of a detector based on Silicon Drift Detectors for gamma-ray spectroscopy and imaging applications / P. Busca, A.D. Butt, C. Fiorini, A. Marone, M. Occhipinti, R. Peloso, R. Quaglia, L. Bombelli, G. Giacomini, C. Piemonte, F. Camera, A. Giaz, B. Million, N. Nelms, B. Shortt. - In: JOURNAL OF INSTRUMENTATION. - ISSN 1748-0221. - 9:5(2014), pp. C05005.1-C05005.12. ((Intervento presentato al 15. convegno International Workshop on Radiation imaging detectors tenutosi a Paris nel 2013 [10.1088/1748-0221/9/05/C05005].

Development of a detector based on Silicon Drift Detectors for gamma-ray spectroscopy and imaging applications

F. Camera;A. Giaz;
2014

Abstract

This work deals with the development of a new gamma detector based on Silicon Drift Detectors (SDDs) to readout large LaBr3:Ce scintillators for gamma-ray spectroscopy and imaging applications. The research is supported by the European Space Agency through the Technology Research Programme (TRP) and by Istituto Nazionale di Fisica Nucleare (INFN) within the Gamma project. The SDDs, produced at Fondazione Bruno Kessler (FBK) semiconductor laboratories, are designed as monolithic arrays of 3 × 3 units, each one of an active area of 8 mm × 8 mm (overall area of 26 mm × 26 mm). The readout electronics and the architecture of the camera are briefly described and then first experimental results coupling the SDD array with a 1'' × 1'' LaBr3:Ce scintillator are reported. An energy resolution of 3% FWHM at 662 keV has been measured at -20°C, better than coupling the same scintillator with a photomultiplier tube. The same scintillator is also used to evaluate position sensitivity with a 1 mm collimated Cs-137 source. The main difficulty in determining the position of the gamma-ray interaction in the crystal is associated to the high thickness/diameter ratio of the crystal (1:1) and the use of reflectors on all lateral and top sides the crystal. This last choice enhances energy resolution but makes imaging capability more challenging because light is spread over all photodetectors. Preliminary results show that the camera is able to detect shifts in the measured signals, when the source is moved with steps of 5 mm. A modified version of the centroid method is finally implemented to evaluate the imaging capability of the system.
Electronic detector readout concepts (solid-state); Front-end electronics for detector readout; Gamma detectors (scintillators, CZT, HPG, HgI etc)
Settore FIS/04 - Fisica Nucleare e Subnucleare
2014
Article (author)
File in questo prodotto:
File Dimensione Formato  
1748-0221_9_05_C05005.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.16 MB
Formato Adobe PDF
1.16 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/248363
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 5
social impact