The objective of this study was to evaluate inert gas rebreathing (IGR) reliability in cardiac output (CO) measurement compared with Fick method and thermodilution. IGR is a noninvasive method for CO measurement; CO by IGR is calculated as pulmonary blood flow plus intrapulmonary shunt. IGR may be ideal for follow-up of patients with pulmonary hypertension (PH), sparing the need of repeated invasive right-sided cardiac catheterization. Right-sided cardiac catheterization with CO measurement by thermodilution, Fick method, and IGR was performed in 125 patients with possible PH by echocardiography. Patients were grouped according to right-sided cardiac catheterization-measured mean pulmonary and wedge pressures: normal pulmonary arterial pressure (n = 20, mean pulmonary arterial pressure = 18 ± 3 mm Hg, pulmonary capillary wedge pressure = 11 ± 5 mm Hg), PH and normal pulmonary capillary wedge pressure (PH-NW, n = 37 mean pulmonary arterial pressure = 42 ± 13 mm Hg, pulmonary capillary wedge pressure = 11 ± 6 mm Hg), and PH and high pulmonary capillary wedge pressure (PH-HW, n = 68, mean pulmonary arterial pressure = 37 ± 9 mm Hg, pulmonary capillary wedge pressure = 24 ± 6 mm Hg). Thermodilution and Fick measurements were comparable. Fick and IGR agreement was observed in normal pulmonary arterial pressure (CO = 4.10 ± 1.14 and 4.08 ± 0.97 L/min, respectively), whereas IGR overestimated Fick in patients with PH-NW and those with PH-HW because of intrapulmonary shunting overestimation in hypoxemic patients. When patients with arterial oxygen saturation (SO2) ≤90% were excluded, IGR and Fick agreement improved in PH-NW (CO = 4.90 ± 1.70 and 4.76 ± 1.35 L/min, respectively) and PH-HW (CO = 4.05 ± 1.04 and 4.10 ± 1.17 L/min, respectively). In hypoxemic patients, we estimated pulmonary shunt as Fick - pulmonary blood flow and calculated shunt as: -0.2423 × arterial SO 2 + 21.373 L/min. In conclusion, IGR is reliable for CO measurement in patients with PH with arterial SO2 >90%. For patients with arterial SO2 ≤90%, a new formula for shunt calculation is proposed.

Noninvasive cardiac output measurement by inert gas rebreathing in suspected pulmonary hypertension / S. Farina, G. Teruzzi, G. Cattadori, C. Ferrari, S. De Martini, M. Bussotti, G. Calligaris, A. Bartorelli, P. Agostoni. - In: THE AMERICAN JOURNAL OF CARDIOLOGY. - ISSN 0002-9149. - 113:3(2014 Feb 01), pp. 546-551. [10.1016/j.amjcard.2013.10.017]

Noninvasive cardiac output measurement by inert gas rebreathing in suspected pulmonary hypertension

S. Farina
Primo
;
G. Cattadori;A. Bartorelli;P. Agostoni
Ultimo
2014

Abstract

The objective of this study was to evaluate inert gas rebreathing (IGR) reliability in cardiac output (CO) measurement compared with Fick method and thermodilution. IGR is a noninvasive method for CO measurement; CO by IGR is calculated as pulmonary blood flow plus intrapulmonary shunt. IGR may be ideal for follow-up of patients with pulmonary hypertension (PH), sparing the need of repeated invasive right-sided cardiac catheterization. Right-sided cardiac catheterization with CO measurement by thermodilution, Fick method, and IGR was performed in 125 patients with possible PH by echocardiography. Patients were grouped according to right-sided cardiac catheterization-measured mean pulmonary and wedge pressures: normal pulmonary arterial pressure (n = 20, mean pulmonary arterial pressure = 18 ± 3 mm Hg, pulmonary capillary wedge pressure = 11 ± 5 mm Hg), PH and normal pulmonary capillary wedge pressure (PH-NW, n = 37 mean pulmonary arterial pressure = 42 ± 13 mm Hg, pulmonary capillary wedge pressure = 11 ± 6 mm Hg), and PH and high pulmonary capillary wedge pressure (PH-HW, n = 68, mean pulmonary arterial pressure = 37 ± 9 mm Hg, pulmonary capillary wedge pressure = 24 ± 6 mm Hg). Thermodilution and Fick measurements were comparable. Fick and IGR agreement was observed in normal pulmonary arterial pressure (CO = 4.10 ± 1.14 and 4.08 ± 0.97 L/min, respectively), whereas IGR overestimated Fick in patients with PH-NW and those with PH-HW because of intrapulmonary shunting overestimation in hypoxemic patients. When patients with arterial oxygen saturation (SO2) ≤90% were excluded, IGR and Fick agreement improved in PH-NW (CO = 4.90 ± 1.70 and 4.76 ± 1.35 L/min, respectively) and PH-HW (CO = 4.05 ± 1.04 and 4.10 ± 1.17 L/min, respectively). In hypoxemic patients, we estimated pulmonary shunt as Fick - pulmonary blood flow and calculated shunt as: -0.2423 × arterial SO 2 + 21.373 L/min. In conclusion, IGR is reliable for CO measurement in patients with PH with arterial SO2 >90%. For patients with arterial SO2 ≤90%, a new formula for shunt calculation is proposed.
Aged ; Breath Tests ; Cardiac Output ; Echocardiography ; Female ; Follow-Up Studies ; Humans ; Hypertension, Pulmonary ; Male ; Middle Aged ; Pulmonary Wedge Pressure ; Reproducibility of Results ; Retrospective Studies ; Severity of Illness Index
Settore MED/11 - Malattie dell'Apparato Cardiovascolare
1-feb-2014
Article (author)
File in questo prodotto:
File Dimensione Formato  
2014 Noninvasive cardiac output measurement by inert gas rebreathing in suspected pulmonary hypertension.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 468.89 kB
Formato Adobe PDF
468.89 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/235624
Citazioni
  • ???jsp.display-item.citation.pmc??? 11
  • Scopus 27
  • ???jsp.display-item.citation.isi??? 25
social impact