Ferrous sulphate gel analysed by relaxation time measurements with NMR imaging is considered a useful dosimeter for 3D determinations of absorbed dose. A protocol for the gel preparation with agarose SeaPlaque that leads to a dosimeter with very high dose sensitivity is described. The dose-response curve slope is about 0.2 s-1 Gy-1 and the G factor turns out to be ~185 ions per 100 eV of absorbed energy. A method for making the measurements and analysing the results that brings about good result reproducibility is suggested. A thorough experimental study of the dependence of the dosimeter response on the elapsed time from preparation to irradiation and from irradiation to NMR measurement has revealed good reproducibility. The above characteristic of the gel system is very interesting, because it shows the possibility of utilizing the dosimeter for absolute dose determinations with satisfactory reliability.

Dose-response curve slope improvement and result reproducibility of ferrous sulphate doped gels analyzed by NMR imaging / G. Gambarini, S. Arrigoni, M.C. Cantone, N. Molho, L. Facchielli, A.E. Sichirollo. - In: PHYSICS IN MEDICINE AND BIOLOGY. - ISSN 0031-9155. - 39:4(1994), pp. 703-717. [10.1088/0031-9155/39/4/004]

Dose-response curve slope improvement and result reproducibility of ferrous sulphate doped gels analyzed by NMR imaging

G. Gambarini
Primo
;
M.C. Cantone;
1994

Abstract

Ferrous sulphate gel analysed by relaxation time measurements with NMR imaging is considered a useful dosimeter for 3D determinations of absorbed dose. A protocol for the gel preparation with agarose SeaPlaque that leads to a dosimeter with very high dose sensitivity is described. The dose-response curve slope is about 0.2 s-1 Gy-1 and the G factor turns out to be ~185 ions per 100 eV of absorbed energy. A method for making the measurements and analysing the results that brings about good result reproducibility is suggested. A thorough experimental study of the dependence of the dosimeter response on the elapsed time from preparation to irradiation and from irradiation to NMR measurement has revealed good reproducibility. The above characteristic of the gel system is very interesting, because it shows the possibility of utilizing the dosimeter for absolute dose determinations with satisfactory reliability.
Settore FIS/07 - Fisica Applicata(Beni Culturali, Ambientali, Biol.e Medicin)
1994
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/225411
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 53
  • ???jsp.display-item.citation.isi??? 47
social impact