The dynamic range of integrated charge-sensitive preamplifiers for germanium detectors is typically limited to ~5 MeV, owing to the intrinsically low available voltage swing of scaled CMOS technologies and to the very high sensitivity needed to achieve energy resolutions of ~0.1% @ 2 MeV. We propose a circuit technique that allows for high-resolution energy measurements, compatible with gamma-spectroscopy standards, also in the case of large input signals yielding a deep saturation of the charge preamplifier. The preamplifier has been designed and simulated in a 5V 0.35 μ m CMOS technology. The idea relies on the fact that the physical information, i.e. the charge released by the germanium crystal, is not destroyed by the saturation of the charge sensing stage. The exceeding charge, that cannot be stored on the feedback capacitance, is temporarily stored on the other capacitances connected at the input node, namely the detector capacitance and the input JFET capacitance. The CMOS circuit comprises a fast-reset device connected at the input node of the circuit and performing the fast de-saturation of the charge sensing stage. A first-order linear relation exists between the input charge and the reset time. By estimating the input charge through the direct measure of the reset time, a substantial increase of the energy measurement range is achieved.

Extending the dynamic range of a charge-preamplifier far beyond its saturation limit : a 0.35 μm CMOS preamplifier for germanium detectors / A. Pullia, F. Zocca - In: 2009 IEEE Nuclear science symposium conference record : October 25–31, 2009 : Orlando, Florida, USA / [a cura di] B. Yu. - Piscataway : IEEE, 2009. - ISBN 9781424439614. - pp. 1919-1923 (( convegno Nuclear science symposium and medical imaging conference tenutosi a Orlando nel 2009.

Extending the dynamic range of a charge-preamplifier far beyond its saturation limit : a 0.35 μm CMOS preamplifier for germanium detectors

A. Pullia
Primo
;
F. Zocca
2009

Abstract

The dynamic range of integrated charge-sensitive preamplifiers for germanium detectors is typically limited to ~5 MeV, owing to the intrinsically low available voltage swing of scaled CMOS technologies and to the very high sensitivity needed to achieve energy resolutions of ~0.1% @ 2 MeV. We propose a circuit technique that allows for high-resolution energy measurements, compatible with gamma-spectroscopy standards, also in the case of large input signals yielding a deep saturation of the charge preamplifier. The preamplifier has been designed and simulated in a 5V 0.35 μ m CMOS technology. The idea relies on the fact that the physical information, i.e. the charge released by the germanium crystal, is not destroyed by the saturation of the charge sensing stage. The exceeding charge, that cannot be stored on the feedback capacitance, is temporarily stored on the other capacitances connected at the input node, namely the detector capacitance and the input JFET capacitance. The CMOS circuit comprises a fast-reset device connected at the input node of the circuit and performing the fast de-saturation of the charge sensing stage. A first-order linear relation exists between the input charge and the reset time. By estimating the input charge through the direct measure of the reset time, a substantial increase of the energy measurement range is achieved.
CMOS preamplifier; Fast-reset preamplifier; Germanium detectors; High-resolution gamma-ray spectroscopy; Low-noise wide-dynamic-range preamplifier
Settore ING-INF/01 - Elettronica
Settore FIS/01 - Fisica Sperimentale
Settore FIS/04 - Fisica Nucleare e Subnucleare
2009
Institute of Electrical and Electronics Engineers (IEEE)
Book Part (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/152458
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
social impact