Background and aims: The monitoring of yearly distributions of HbA2 measured has been indicated as a reliable indicator of worldwide standardization. Materials and methods: Measurements/year of HbA2 have been collected over three consecutive years in 15 Italian laboratories each using the same analytical method over three years period. HbA2 distributions, cleaned of replicated measurements, were compared by the overlapping area of the raw probability density functions expressed by coefficient eta (η), and by comparing the reference intervals for the central part of each distribution estimated by the indirect method refineR using the R package "refineR". Results: According to the overlapping areas analysis the distributions/year of the data provided by 4 centers able to perform at least 1000 measurements/year were similar in 2 consecutive years. Moreover, the reference intervals provided by 2 centers using the same analytical methods in two separate locations over the three consecutive years, were very similar. The highest overlap (99.7 %) was observed in one center over two consecutive years. The overlapping areas were very high (93.6-95.7%) in 8 out of 9 inter-comparisons. Conclusion: Despite the limitations of this study the yearly distribution of the HbA2 measured in various centers appears a reliable tool to test HbA2 standardization over different centers using different analytical methods.

Analysis of annual distributions of hemoglobin A2 values as a method to test for HbA2 standardization / M. Vidali, R. Paleari, F. Ceriotti, S. Bernardini, I. Giambini, D. Brugnoni, L. Trainini, C. Carru, F. Porcu, M. Carta, D. Giavarina, M. Ciaccio, B. Lo Sasso, G. Corso, C. Paolillo, R. Dorizzi, M. Rosetti, M. Fiorini, M. Bombara, M. Grosso, M. Giuliano, M. Locatelli, A. Lucci, F. Mencarelli, N. Scattolo, M. Strollo, A. Mosca. - In: CLINICA CHIMICA ACTA. - ISSN 0009-8981. - 550:(2023 Oct 01), pp. 117564.1-117564.10. [10.1016/j.cca.2023.117564]

Analysis of annual distributions of hemoglobin A2 values as a method to test for HbA2 standardization

R. Paleari
Secondo
;
A. Mosca
Ultimo
2023

Abstract

Background and aims: The monitoring of yearly distributions of HbA2 measured has been indicated as a reliable indicator of worldwide standardization. Materials and methods: Measurements/year of HbA2 have been collected over three consecutive years in 15 Italian laboratories each using the same analytical method over three years period. HbA2 distributions, cleaned of replicated measurements, were compared by the overlapping area of the raw probability density functions expressed by coefficient eta (η), and by comparing the reference intervals for the central part of each distribution estimated by the indirect method refineR using the R package "refineR". Results: According to the overlapping areas analysis the distributions/year of the data provided by 4 centers able to perform at least 1000 measurements/year were similar in 2 consecutive years. Moreover, the reference intervals provided by 2 centers using the same analytical methods in two separate locations over the three consecutive years, were very similar. The highest overlap (99.7 %) was observed in one center over two consecutive years. The overlapping areas were very high (93.6-95.7%) in 8 out of 9 inter-comparisons. Conclusion: Despite the limitations of this study the yearly distribution of the HbA2 measured in various centers appears a reliable tool to test HbA2 standardization over different centers using different analytical methods.
No
English
Harmonization; Hemoglobin A(2); Reference intervals; Standardization; β-thalassemia
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
Articolo
Esperti anonimi
Pubblicazione scientifica
1-ott-2023
Elsevier
550
117564
1
10
10
Pubblicato
Periodico con rilevanza internazionale
pubmed
scopus
crossref
Aderisco
info:eu-repo/semantics/article
Analysis of annual distributions of hemoglobin A2 values as a method to test for HbA2 standardization / M. Vidali, R. Paleari, F. Ceriotti, S. Bernardini, I. Giambini, D. Brugnoni, L. Trainini, C. Carru, F. Porcu, M. Carta, D. Giavarina, M. Ciaccio, B. Lo Sasso, G. Corso, C. Paolillo, R. Dorizzi, M. Rosetti, M. Fiorini, M. Bombara, M. Grosso, M. Giuliano, M. Locatelli, A. Lucci, F. Mencarelli, N. Scattolo, M. Strollo, A. Mosca. - In: CLINICA CHIMICA ACTA. - ISSN 0009-8981. - 550:(2023 Oct 01), pp. 117564.1-117564.10. [10.1016/j.cca.2023.117564]
reserved
Prodotti della ricerca::01 - Articolo su periodico
27
262
Article (author)
Periodico con Impact Factor
M. Vidali, R. Paleari, F. Ceriotti, S. Bernardini, I. Giambini, D. Brugnoni, L. Trainini, C. Carru, F. Porcu, M. Carta, D. Giavarina, M. Ciaccio, B. L...espandi
File in questo prodotto:
File Dimensione Formato  
2023_distribuzioni A2.pdf

accesso riservato

Descrizione: Article
Tipologia: Publisher's version/PDF
Dimensione 4.83 MB
Formato Adobe PDF
4.83 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/1018949
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
  • OpenAlex ND
social impact