OBJECTIVES: To test the in vivo effects of an ultrasonic dental scaler on various implanted cardioverter defibrillator (ICD) models. METHODS: 12 consecutive patients with ICDs had continuous both electrocardiogram monitoring and device interrogation to detect interferences during the use of an ultrasonic dental scaler. RESULTS: No interferences were detected by any ICD. Evaluation of the electrocardiograms for each patient failed to show any abnormalities in pacing during testing. CONCLUSION: The results of this study suggest that the routinary clinic use of piezoelectric dental scalers do not interfere with the functioning of any of the tested ICDs. CLINICAL SIGNIFICANCE: Ultrasonic dental scalers have been suspected of electromagnetic interference (EMI) with the normal functioning of ICDs and the use of this type of equipment for patients with these devices has been controversial. This is the first in vivo study to investigate EMI of ICD activity during the operation with ultrasonic dental scaler.

Do ultrasonic dental scalers interfere with implantable cardioverter defibrillators? An in vivo investigation / C. Maiorana, G.B. Grossi, R.A. Garramone, R. Manfredini, F. Santoro. - In: JOURNAL OF DENTISTRY. - ISSN 0300-5712. - 41:11(2013 Nov), pp. 955-959.

Do ultrasonic dental scalers interfere with implantable cardioverter defibrillators? An in vivo investigation

C. Maiorana
Primo
;
G.B. Grossi
Secondo
;
R.A. Garramone;F. Santoro
Ultimo
2013

Abstract

OBJECTIVES: To test the in vivo effects of an ultrasonic dental scaler on various implanted cardioverter defibrillator (ICD) models. METHODS: 12 consecutive patients with ICDs had continuous both electrocardiogram monitoring and device interrogation to detect interferences during the use of an ultrasonic dental scaler. RESULTS: No interferences were detected by any ICD. Evaluation of the electrocardiograms for each patient failed to show any abnormalities in pacing during testing. CONCLUSION: The results of this study suggest that the routinary clinic use of piezoelectric dental scalers do not interfere with the functioning of any of the tested ICDs. CLINICAL SIGNIFICANCE: Ultrasonic dental scalers have been suspected of electromagnetic interference (EMI) with the normal functioning of ICDs and the use of this type of equipment for patients with these devices has been controversial. This is the first in vivo study to investigate EMI of ICD activity during the operation with ultrasonic dental scaler.
electromagnetic interference ; implantable cardioverter defibrillator ; ultrasonic dental scaler
Settore MED/28 - Malattie Odontostomatologiche
nov-2013
Article (author)
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S030057121300208X-main.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 1.66 MB
Formato Adobe PDF
1.66 MB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/234989
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 12
  • ???jsp.display-item.citation.isi??? 12
social impact