The efficiency and accuracy of different methods for quality control of radiopharmaceutical preparations for diagnostic purpose were studied. The radiochemical purity of 99mTc Tetrafosmin, 99mTc Exametazime, 99mTc Sestamibi and 99mTc Oxidronate was evaluated by different thin layer chromatography systems, followed by cutting of the strips into two or three sections and by the measurement of radioactivity distribution by dose calibrator or gamma counter. In addition, to confirm the accuracy of these routine procedures, the strips were cut into a number of micro-sections (14–25) and each of them evaluated by the gamma counter. The three tested procedures gave similar results and revealed a good and comparable accuracy. The radioactivity measurement with the dose calibrator remains the most practicable because of the rapidity of execution.

Evaluation of different counting methods for use in radiochemical purity testing procedures for (99m)Tc-labelled radiopharmaceuticals / C. Maioli, A. Bestetti, F. Milani, G.P. Cornalba, L. Tagliabue, D. Di Benedetto, I. Rognoni, G.L. Tarolo, R. Paroni. - In: APPLIED RADIATION AND ISOTOPES. - ISSN 0969-8043. - 66:4(2008 Apr), pp. 556-559. [10.1016/j.apradiso.2007.07.021]

Evaluation of different counting methods for use in radiochemical purity testing procedures for (99m)Tc-labelled radiopharmaceuticals

C. Maioli
Primo
;
A. Bestetti
Secondo
;
F. Milani;G.P. Cornalba;G.L. Tarolo
Penultimo
;
R. Paroni
Ultimo
2008

Abstract

The efficiency and accuracy of different methods for quality control of radiopharmaceutical preparations for diagnostic purpose were studied. The radiochemical purity of 99mTc Tetrafosmin, 99mTc Exametazime, 99mTc Sestamibi and 99mTc Oxidronate was evaluated by different thin layer chromatography systems, followed by cutting of the strips into two or three sections and by the measurement of radioactivity distribution by dose calibrator or gamma counter. In addition, to confirm the accuracy of these routine procedures, the strips were cut into a number of micro-sections (14–25) and each of them evaluated by the gamma counter. The three tested procedures gave similar results and revealed a good and comparable accuracy. The radioactivity measurement with the dose calibrator remains the most practicable because of the rapidity of execution.
Radiochemical purity ; Radiopharmaceuticals ; Quality control ; ITLC-SG chromatography
Settore BIO/10 - Biochimica
Settore MED/36 - Diagnostica per Immagini e Radioterapia
apr-2008
Article (author)
File in questo prodotto:
File Dimensione Formato  
96-Evaluation of different counting methods for use in radiochemical Appl Rad Inter 2008.pdf

accesso aperto

Tipologia: Publisher's version/PDF
Dimensione 295.87 kB
Formato Adobe PDF
295.87 kB Adobe PDF Visualizza/Apri
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/35044
Citazioni
  • ???jsp.display-item.citation.pmc??? 1
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 6
social impact