PURPOSE: To check the interobserver agreement between radiation oncologists and therapists (RTT) using an on- and off-line cone-beam computer tomography (CBCT) protocol for setup verification in the radiotherapy of prostate cancer. MATERIALS AND METHODS: The CBCT data from six prostate cancer patients treated with hypofractionated intensity-modulated radiotherapy (IMRT) were independently reviewed off-line by four observers (one radiation oncologist, one junior and two senior RTTs) and benchmarked with on-line CBCT positioning performed by a radiation oncologist immediately prior to treatment. CBCT positioning was based on manual soft-tissue registration. Agreement between observers was evaluated using weighted Cohen's kappa statistics. RESULTS: In total, 152 CBCT-based prostate positioning procedures were reviewed by each observer. The mean (± standard deviation) of the differences between off- and on-line CBCT-simCT registration translations along the three directions (antero-posterior, latero-lateral and cranio-caudal) and rotation around the antero-posterior axis were - 0.7 (3.6) mm, 1.9 (2.7) mm, 0.9 (3.6) mm and - 1.8 (5.0) degrees, respectively. Satisfactory interobserver agreement was found, being substantial (weighted kappa > 0.6) in 10 of 16 comparisons and moderate (0.41-0.60) in the remaining six comparisons. CONCLUSIONS: CBCT interpretation performed by RTTs is comparable to that of radiation oncologists. Our study might be helpful in the quality assurance of radiotherapy and the optimization of competencies. Further investigation should include larger sample sizes, a greater number of observers and validated methodology in order to assess interobserver variability and its impact on high-precision prostate cancer IGRT. In the future, it should enable the wider implementation of complex and evolving radiotherapy technologies.

Prostate positioning using cone-beam computer tomography based on manual soft-tissue registration : interobserver agreement between radiation oncologists and therapists / B.A. Jereczek-Fossa, C. Pobbiati, L. Santoro, C. Fodor, P. Fanti, S. Vigorito, G. Baroni, D. Zerini, O. De Cobelli, R. Orecchia. - In: STRAHLENTHERAPIE UND ONKOLOGIE. - ISSN 0179-7158. - 190:1(2014), pp. 81-87.

Prostate positioning using cone-beam computer tomography based on manual soft-tissue registration : interobserver agreement between radiation oncologists and therapists

B.A. Jereczek-Fossa
Primo
;
O. De Cobelli
Penultimo
;
R. Orecchia
Ultimo
2014

Abstract

PURPOSE: To check the interobserver agreement between radiation oncologists and therapists (RTT) using an on- and off-line cone-beam computer tomography (CBCT) protocol for setup verification in the radiotherapy of prostate cancer. MATERIALS AND METHODS: The CBCT data from six prostate cancer patients treated with hypofractionated intensity-modulated radiotherapy (IMRT) were independently reviewed off-line by four observers (one radiation oncologist, one junior and two senior RTTs) and benchmarked with on-line CBCT positioning performed by a radiation oncologist immediately prior to treatment. CBCT positioning was based on manual soft-tissue registration. Agreement between observers was evaluated using weighted Cohen's kappa statistics. RESULTS: In total, 152 CBCT-based prostate positioning procedures were reviewed by each observer. The mean (± standard deviation) of the differences between off- and on-line CBCT-simCT registration translations along the three directions (antero-posterior, latero-lateral and cranio-caudal) and rotation around the antero-posterior axis were - 0.7 (3.6) mm, 1.9 (2.7) mm, 0.9 (3.6) mm and - 1.8 (5.0) degrees, respectively. Satisfactory interobserver agreement was found, being substantial (weighted kappa > 0.6) in 10 of 16 comparisons and moderate (0.41-0.60) in the remaining six comparisons. CONCLUSIONS: CBCT interpretation performed by RTTs is comparable to that of radiation oncologists. Our study might be helpful in the quality assurance of radiotherapy and the optimization of competencies. Further investigation should include larger sample sizes, a greater number of observers and validated methodology in order to assess interobserver variability and its impact on high-precision prostate cancer IGRT. In the future, it should enable the wider implementation of complex and evolving radiotherapy technologies.
CBCT; interobserver agreement; prostate cancer radiotherapy; radiation oncologist; radiation therapist
Settore MED/36 - Diagnostica per Immagini e Radioterapia
Settore MED/24 - Urologia
2014
Article (author)
File in questo prodotto:
File Dimensione Formato  
art_10.1007_s00066-013-0387-1.pdf

accesso riservato

Tipologia: Publisher's version/PDF
Dimensione 436.2 kB
Formato Adobe PDF
436.2 kB Adobe PDF   Visualizza/Apri   Richiedi una copia
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/224365
Citazioni
  • ???jsp.display-item.citation.pmc??? 8
  • Scopus 13
  • ???jsp.display-item.citation.isi??? 11
social impact