Aims: The recommended method for the international sensitivity index (ISI) calibration of whole blood point of care testing (POCT) prothrombin time (PT) systems was originally described by Tripodi et al in 1993 but is too complex and demanding. The present European Concerted Action on Anticoagulation (ECAA) study aimed to assess the reliability of simpler ISI calibration using lyophilised plasma samples. Methods: ISI calibrations using three different types of ECAA lyophilised plasma samples (artificially depleted, individual, and pooled coumarin) were compared with whole blood calibrations on CoaguChek Mini and TAS PT-NC POCT monitors at 10 centres. Results: With CoaguChek Mini systems, lyophilised coumarin plasma samples (both single donation and pooled) gave ISI and international normalised ratio (INR) values comparable to whole blood. With artificially depleted plasma, ISI and INR values were too high. With TAS PT-NC systems, all three types of lyophilised plasma samples gave inaccurate ISI and unreliable INR results, similar to previous ECAA findings with fresh plasma calibrations. Conclusions: With CoaguChek Mini systems, ISI calibration can be simplified by the use of ECAA lyophilised plasma samples from coumarin treated patients. Further study is needed to devise a simpler calibration method for the TAS PT-NC system.

European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors / L. Poller, M. Keown, N. Chauhan, A.M.H.P. van den Besselaar, A. Tripodi, J. Jespersen, C. Shiach. - In: JOURNAL OF CLINICAL PATHOLOGY. - ISSN 0021-9746. - 56:2(2003), pp. 114-119.

European Concerted Action on Anticoagulation (ECAA). An assessment of lyophilised plasmas for ISI calibration of CoaguChek and TAS whole blood prothrombin time monitors

A. Tripodi;
2003

Abstract

Aims: The recommended method for the international sensitivity index (ISI) calibration of whole blood point of care testing (POCT) prothrombin time (PT) systems was originally described by Tripodi et al in 1993 but is too complex and demanding. The present European Concerted Action on Anticoagulation (ECAA) study aimed to assess the reliability of simpler ISI calibration using lyophilised plasma samples. Methods: ISI calibrations using three different types of ECAA lyophilised plasma samples (artificially depleted, individual, and pooled coumarin) were compared with whole blood calibrations on CoaguChek Mini and TAS PT-NC POCT monitors at 10 centres. Results: With CoaguChek Mini systems, lyophilised coumarin plasma samples (both single donation and pooled) gave ISI and international normalised ratio (INR) values comparable to whole blood. With artificially depleted plasma, ISI and INR values were too high. With TAS PT-NC systems, all three types of lyophilised plasma samples gave inaccurate ISI and unreliable INR results, similar to previous ECAA findings with fresh plasma calibrations. Conclusions: With CoaguChek Mini systems, ISI calibration can be simplified by the use of ECAA lyophilised plasma samples from coumarin treated patients. Further study is needed to devise a simpler calibration method for the TAS PT-NC system.
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
Settore MED/09 - Medicina Interna
Settore MED/15 - Malattie del Sangue
2003
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/189989
Citazioni
  • ???jsp.display-item.citation.pmc??? 3
  • Scopus 15
  • ???jsp.display-item.citation.isi??? 13
social impact