The same frozen plasma samples, with normal and high contents of fibrinogen, were examined in four laboratories. The methods were calibrated with an internal standard whose fibrinogen content was determined gravimetrically. Both the Von Clauss and the RID methods were reliable, accurate, and precise, if adequate calibration was used. The PT-derived method was highly reproducible, but had some problems with accuracy. We demonstrate that an adequate calibration procedure is indispensable for reliable fibrinogen measurements whatever method is used. Because neither the calibration procedures proposed by the manufacturers nor the use of lyophilized commercial plasmas is adequate for this purpose, we urge that an international standard for fibrinogen measurement be promptly established.

Fibrinogen Assays: A collaborative Study of six different methods / G. Palareti, M. Maccaferri, C. Manotti, A. Tripodi, V. Chantarangkul, F. Rodeghiero, M. Ruggeri, P.M. Mannucci. - In: CLINICAL CHEMISTRY. - ISSN 0009-9147. - 37:5(1991), pp. 714-719.

Fibrinogen Assays: A collaborative Study of six different methods

A. Tripodi;
1991

Abstract

The same frozen plasma samples, with normal and high contents of fibrinogen, were examined in four laboratories. The methods were calibrated with an internal standard whose fibrinogen content was determined gravimetrically. Both the Von Clauss and the RID methods were reliable, accurate, and precise, if adequate calibration was used. The PT-derived method was highly reproducible, but had some problems with accuracy. We demonstrate that an adequate calibration procedure is indispensable for reliable fibrinogen measurements whatever method is used. Because neither the calibration procedures proposed by the manufacturers nor the use of lyophilized commercial plasmas is adequate for this purpose, we urge that an international standard for fibrinogen measurement be promptly established.
Calibration; Coagulation; Intermethod comparison
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
Settore MED/09 - Medicina Interna
Settore MED/15 - Malattie del Sangue
1991
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/189424
Citazioni
  • ???jsp.display-item.citation.pmc??? 4
  • Scopus 54
  • ???jsp.display-item.citation.isi??? ND
social impact