The international normalized ratio (INR) on two widely used point-of-care test (POCT) prothrombin time (PT) monitors (CoaguChek Mini and TAS PT-NC) differed considerably and also differed from the 'true' INR obtained on the same samples using a manual PT and the same species thromboplastin international reference preparation. Agreement between the displayed INR and difference from 'true' INR has been reassessed following an independent international sensitivity index (ISI) calibration of the two systems. The displayed INRs taken at seven centres were compared with 'true' INRs from the same blood donations and INRs based on the resulting ISI. The overall difference between the displayed INRs on the two monitor systems was reduced from 21.0% to 3.5%. The overall difference in mean INR of system A from the 'true' INR was reduced from 19.0% to 9.5% and of system B from 6.8% to 0.3%, but individual centre's results still showed considerable mean INR variability. Differences between overall displayed INR with the two monitor systems have been reduced by an independent multicentre calibration, and agreement with 'true' INR on the same blood samples improved. However, marked variability in mean INR at individual centres remained after ISI correction, which demonstrates the need for external quality control of individual POCT whole-blood PT monitors.

European concerted action on anticoagulation. Correction of displayed international normalised ratio on two point-of-care test whole-blood prothrombin time monitors (CoaguChek Mini and TAS PT-NC) by independent international sensitivity index calibration / L. Poller, M. Keown, N. Chauhan, A.M.H.P. van den Besselaar, A. Tripodi, C. Shiach, J. Jespersen. - In: BRITISH JOURNAL OF HAEMATOLOGY. - ISSN 0007-1048. - 122:6(2003), pp. 944-949.

European concerted action on anticoagulation. Correction of displayed international normalised ratio on two point-of-care test whole-blood prothrombin time monitors (CoaguChek Mini and TAS PT-NC) by independent international sensitivity index calibration

A. Tripodi;
2003

Abstract

The international normalized ratio (INR) on two widely used point-of-care test (POCT) prothrombin time (PT) monitors (CoaguChek Mini and TAS PT-NC) differed considerably and also differed from the 'true' INR obtained on the same samples using a manual PT and the same species thromboplastin international reference preparation. Agreement between the displayed INR and difference from 'true' INR has been reassessed following an independent international sensitivity index (ISI) calibration of the two systems. The displayed INRs taken at seven centres were compared with 'true' INRs from the same blood donations and INRs based on the resulting ISI. The overall difference between the displayed INRs on the two monitor systems was reduced from 21.0% to 3.5%. The overall difference in mean INR of system A from the 'true' INR was reduced from 19.0% to 9.5% and of system B from 6.8% to 0.3%, but individual centre's results still showed considerable mean INR variability. Differences between overall displayed INR with the two monitor systems have been reduced by an independent multicentre calibration, and agreement with 'true' INR on the same blood samples improved. However, marked variability in mean INR at individual centres remained after ISI correction, which demonstrates the need for external quality control of individual POCT whole-blood PT monitors.
Settore BIO/12 - Biochimica Clinica e Biologia Molecolare Clinica
Settore MED/09 - Medicina Interna
Settore MED/15 - Malattie del Sangue
2003
Article (author)
File in questo prodotto:
Non ci sono file associati a questo prodotto.
Pubblicazioni consigliate

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2434/189158
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 8
social impact