Two ref. electrodes, Hg(I) sulfate and Pb(I) sulfate, with incorporated Li2SO4 salt bridge, are proposed as alternatives for diffuse electrodes of calomel and silver chloride, in potentiometric monitoring such as pH-metry and pIonometry. With a Li2SO4 salt bridge of 2m, the operating potential of the two ref. electrodes is, resp., EREF[Hg4SO4] = 0.6326 ± 0.0008 V and EREF[PbSO4] = -0.3327 ± 0.0008 V, at 25°.

Alternative reference electrodes / P.R. Mussini, T. Mussini. - In: LABORATORIO 2000. - ISSN 1120-8376. - 12:1(1998), pp. 58-59.

Alternative reference electrodes

P.R. Mussini
Primo
;
T. Mussini
Ultimo
1998

Abstract

Two ref. electrodes, Hg(I) sulfate and Pb(I) sulfate, with incorporated Li2SO4 salt bridge, are proposed as alternatives for diffuse electrodes of calomel and silver chloride, in potentiometric monitoring such as pH-metry and pIonometry. With a Li2SO4 salt bridge of 2m, the operating potential of the two ref. electrodes is, resp., EREF[Hg4SO4] = 0.6326 ± 0.0008 V and EREF[PbSO4] = -0.3327 ± 0.0008 V, at 25°.